Biomechanical characterisation of a degradable magnesium-based (MgCa0.8) screw
- 31 December 2011
- journal article
- Published by Springer Nature in Journal of Materials Science: Materials in Medicine
- Vol. 23 (3) , 649-655
Abstract
Magnesium alloys have been in the focus of research in recent years as degradable biomaterial. The purpose of this study was the biomechanical characterisation of MgCa0.8-screws. The maximum pull out force of screws was determined in a synthetic bone without corrosion and after fixed intervals of corrosion: 24, 48, 72 and 96 h. This in vitro study has been carried out with Hank’s solution with a flow rate corresponding to the blood flow in natural bone. A maximum pull out force (Fmax) of 201.5 ± 9.3 N was measured without corrosion. The biomechanical parameter decreased by 30% after 96 h in corrosive medium compared to the non-corrosion group. A maximum load capacity of 28 ± 7.6 N/h was determined. Our biomechanical data suggests that this biodegradable screw provides a promising bone-screw-fixation and has great potential for medical application.Keywords
This publication has 29 references indexed in Scilit:
- In‐vitro dissolution of magnesium–calcium binary alloys: Clarifying the unique role of calcium additions in bioresorbable magnesium implant alloysJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2010
- Biomedical Implants: Corrosion and its Prevention - A ReviewRecent Patents on Corrosion Science, 2010
- Degradable biomaterials based on magnesium corrosionCurrent Opinion in Solid State and Materials Science, 2008
- Progress and Challenge for Magnesium Alloys as BiomaterialsAdvanced Engineering Materials, 2008
- Stability of connected mini-implants and miniplates for skeletal anchorage in orthodonticsEuropean Journal of Orthodontics, 2008
- In vivo corrosion of four magnesium alloys and the associated bone responseBiomaterials, 2005
- Understanding Magnesium Corrosion—A Framework for Improved Alloy PerformanceAdvanced Engineering Materials, 2003
- Mechanical performance of standard and cannulated 4.0‐mm cancellous bone screwsJournal of Orthopaedic Research, 2000
- Factors Affecting the Pullout Strength of Cancellous Bone ScrewsJournal of Biomechanical Engineering, 1996
- Mechanical characterization of biodegradable implantsClinical Materials, 1992