HIFU Power Monitoring Using Combined Instantaneous Current and Voltage Measurement
- 12 September 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 67 (2) , 239-247
Abstract
During high-intensity focused ultrasound (HIFU) therapy, it is important that the electrical power delivered to the transducer is monitored to avoid underexposure or overexposure, ensure patient safety, and to protect the transducer itself. Due to ease of measurement, the transducer's potential difference may be as an indicator of power delivery. However, even when a transducer's complex impedance is well characterized at small amplitudes and matching networks are used, voltage-only (VO) monitoring cannot account for the presence of drive waveform distortion, changes to the acoustic path, or damage to the transducer. In this study, combined current and voltage (CCV) is proposed as a magnetic resonance imaging (MRI)-compatible, miniature alternative to bidirectional power couplers, which is compatible with switched amplifiers. For CCV power measurement, current probe data were multiplied by the voltage waveform and integrated in the frequency domain. Transducer efficiency was taken into account to predict acoustic power. The technique was validated with a radiation force balance (RFB). When using a typical HIFU transducer and amplifier, VO predictions and acoustic power had a maximum difference of 20%. However, under the same conditions, CCV only had a maximum difference of 5%. The technique was applied to several lesioning experiments and it was shown that when VO was used as a control between two amplifiers, there was up to a 38% difference in lesion area. This greatly reduced to a maximum of 5% once CCV was used instead. These results demonstrate that CCV can accurately predict real-time electrical power delivery, leading to safer HIFU treatments.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/K029835/1)
- National Heart, Lung, and Blood Institute (1 U01 HL121838-01)
- Wellcome Trust (105615/Z/14/Z)
- Royal Society (RG170324)
This publication has 51 references indexed in Scilit:
- Noninvasive Treatment of Deep Venous Thrombosis Using Pulsed Ultrasound Cavitation Therapy (Histotripsy) in a Porcine ModelJournal of Vascular and Interventional Radiology, 2011
- Real-Time Monitoring of High-Intensity Focused Ultrasound Lesion Formation Using Acousto-Optic SensingUltrasound in Medicine & Biology, 2011
- Instantaneous input electrical power measurements of HITU transducerJournal of Physics: Conference Series, 2011
- Driving Circuitry for Focused Ultrasound Noninvasive Surgery and Drug Delivery ApplicationsSensors, 2011
- Acoustic characterization of high intensity focused ultrasound fields: A combined measurement and modeling approachThe Journal of the Acoustical Society of America, 2008
- High-intensity focused ultrasound in the treatment of solid tumoursNature Reviews Cancer, 2005
- 500‐element ultrasound phased array system for noninvasive focal surgery of the brain: A preliminary rabbit study with ex vivo human skullsMagnetic Resonance in Medicine, 2004
- High intensity focused ultrasound: surgery of the future?British Journal of Radiology, 2003
- Design and experimental evaluation of an intracavitary ultrasound phased array system for hyperthermiaIEEE Transactions on Biomedical Engineering, 1994
- MR-Guided Focused Ultrasound SurgeryJournal of Computer Assisted Tomography, 1992