A 32-Channel Neural-Recording Chip Achieving 117dB Intrinsic-CMRR and 100dB PSRR by CM-Tracking-Dynamic-Power-Rail and CM-Canceling-in-Idle-Phase Techniques
- 10 April 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. PP (19324545) , 1-15
Abstract
Multi-channel neural recording enables simultaneous monitoring of neuronal activities across multiple brain regions, while the in-vivo common-mode interference (CMI) significantly degrades the signal quality of implantable neural-recording chips. For a multi-channel neural-recording chip, the total common-mode rejection ratio (T-CMRR) of the analog front-end (AFE) is limited by the input imbalance between the signal electrode and the shared reference electrode, as well as the intrinsic CMRR (I-CMRR) of its circuits. The traditional common-mode replication (CM-REP) technique is only applicable to single-channel systems such as ECG monitoring devices. In addition, conventional pre-amplifier and frequency-controlled differential regulator (FCDR) techniques suffer from gain mismatch and high power consumption, respectively. To address these issues, this work presents a 32-channel neural-recording chip fabricated in a 65 nm CMOS process, which effectively suppresses the CMI in two operational modes: 1) In high-gain mode, the proposed CM-tracking-dynamic-power-rail (CM-TDPR) instrumentation amplifier (IA) achieves 50 GΩ CM input impedance, 117 dB I-CMRR, and 100 dB power supply rejection ratio (PSRR), resulting in a T-CMRR of 87 dB; 2) In low-gain mode, a CM-canceling-in-idle-phase (CM-CIP) technique is proposed to increase the I-CMRR to 102 dB and match the signal-reference input impedance, thereby achieving a 95 dB T-CMRR. In-vivo experiments were conducted on a Sprague-Dawley rat, successfully validating the CMRR performance of the proposed chip.Keywords
Funding Information
- National Key Research and Development Program of China (2023YFF1203600)
- Welfare Ethics Review Committee of the Experimental Animal Center of Zhejiang University (ZJU20190116)
This publication has 25 references indexed in Scilit:
- A 4.6μW 3.3-NEF Biopotential Amplifier with 133VPP Common-Mode Interference Tolerance and 102dB Total Common-Mode Rejection Ratio for Two-Electrode Recording System2025
- 31.4 A 128-Channel 2mmx2mm Battery-Free Neural Dielet Merging Simultaneous Multi-Channel Transmission Through Multi-Carrier Orthogonal Backscatter2023
- A Miniaturized Wireless Neural Implant with Body-Coupled Data Transmission and Power Delivery for Freely Behaving Animals2022
- A Time-Division Multiplexed 8-Channel Non-Contact ECG Recording IC with a Common-Mode Interference Tolerance of $20\mathrm{V}_{\text{PP}}$2022
- 34.6 EEG Dust: A BCC-Based Wireless Concurrent Recording/Transmitting Concentric Electrode2020
- A 0.8V Chopper Amplifier with 600mVpp Tolerance to Power-Line Interference for Neural Signal Acquisition2019
- A 110dB-CMRR 100dB-PSRR multi-channel neural-recording amplifier system using differentially regulated rejection ratio enhancement in 0.18μm CMOS2018
- A 0.45V 100-channel neural-recording IC with sub-µW/channel consumption in 0.18µm CMOS2013
- A 0.013mm2 5μW DC-coupled neural signal acquisition IC with 0.5V supply2011
- Large-scale recording of neuronal ensemblesNature Neuroscience, 2004