Comprehensive analysis of proton range uncertainties related to stopping-power-ratio estimation using dual-energy CT imaging
- 9 August 2017
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (17) , 7056-7074
Abstract
The dual-energy CT-based (DECT) approach holds promise in reducing the overall uncertainty in proton stopping-power-ratio (SPR) estimation as compared to the conventional stoichiometric calibration approach. The objective of this study was to analyze the factors contributing to uncertainty in SPR estimation using the DECT-based approach and to derive a comprehensive estimate of the range uncertainty associated with SPR estimation in treatment planning. Two state-of-the-art DECT-based methods were selected and implemented on a Siemens SOMATOM Force DECT scanner. The uncertainties were first divided into five independent categories. The uncertainty associated with each category was estimated for lung, soft and bone tissues separately. A single composite uncertainty estimate was eventually determined for three tumor sites (lung, prostate and head-and-neck) by weighting the relative proportion of each tissue group for that specific site. The uncertainties associated with the two selected DECT methods were found to be similar, therefore the following results applied to both methods. The overall uncertainty (1 σ ) in SPR estimation with the DECT-based approach was estimated to be 3.8%, 1.2% and 2.0% for lung, soft and bone tissues, respectively. The dominant factor contributing to uncertainty in the DECT approach was the imaging uncertainties, followed by the DECT modeling uncertainties. Our study showed that the DECT approach can reduce the overall range uncertainty to approximately 2.2% (2 σ ) in clinical scenarios, in contrast to the previously reported 1%.Keywords
Funding Information
- National Natural Science Foundation of China (81571771)
- Cancer Prevention and Research Institute of Texas (RP160661)
- Ministry of Science and Technology of the People's Republic of China (2015BAI01B10)
- National Institutes of Health (P20CA183639-01A1)
This publication has 31 references indexed in Scilit:
- The potential of dual-energy CT to reduce proton beam range uncertaintiesMedical Physics, 2017
- A linear, separable two-parameter model for dual energy CT imaging of proton stopping power computationMedical Physics, 2016
- Comparison of proton therapy treatment planning for head tumors with a pencil beam algorithm on dual and single energy CT imagesMedical Physics, 2016
- A simulation study on proton computed tomography (CT) stopping power accuracy using dual energy CT scans as benchmarkActa Oncologica, 2015
- Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogatesPhysics in Medicine & Biology, 2014
- Promise and Pitfalls of Heavy-Particle TherapyJournal of Clinical Oncology, 2014
- A stoichiometric calibration method for dual energy computed tomographyPhysics in Medicine & Biology, 2014
- Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogatesPhysics in Medicine & Biology, 2013
- Physics Controversies in Proton TherapySeminars in Radiation Oncology, 2013
- The influence of CT image noise on proton range calculation in radiotherapy planningPhysics in Medicine & Biology, 2010