Asynchronous calibration of a CT scanner for bone mineral density estimation: sources of error and correction
Authors:
- Alice DudleADAlice Dudle
- Michael IthMIMichael Ith
- Rainer EgliRERainer Egli
- Johannes HeverhagenJHJohannes Heverhagen
- Yvan GuglerYGYvan Gugler
- Christina WappCWChristina Wapp
- Daniela A FrauchigerDFDaniela A Frauchiger
- Kurt LippunerKLKurt Lippuner
- Christian JackowskiCJChristian Jackowski
- Philippe ZyssetPZPhilippe Zysset
Open Access
- 23 July 2024
- journal article
- research article
- Published by Oxford University Press (OUP) in JBMR Plus
- Vol. 8 (9) , ziae096
Abstract
The estimation of BMD with CT scans requires a calibration method, usually based on a phantom. In asynchronous calibration, the phantom is scanned separately from the patient. A standardized acquisition protocol must be used to avoid variations between patient and phantom. However, variations can still be induced, for example, by temporal fluctuations or patient characteristics. Based on the further use of 739 forensic and 111 clinical CT scans, this study uses the proximal femur BMD value (“total hip”) to assess asynchronous calibration accuracy, using in-scan calibration as ground truth. It identifies the parameters affecting the calibration accuracy and quantifies their impact. For time interval and table height, the impact was measured by calibrating the CT scan twice (once using the phantom scan with closest acquisition parameters and once using a phantom scan with standard values) and comparing the calibration accuracy. For other parameters such as body weight, the impact was measured by computing a linear regression between parameter values and calibration accuracy. Finally, this study proposes correction methods to reduce the effect of these parameters and improve the calibration accuracy. The BMD error of the asynchronous calibration, using the phantom scan with the closest acquisition parameters, was −1.2 ± 1.7% for the forensic and − 1.6 ± 3.5% for the clinical dataset. Among the parameters studied, time interval and body weight were identified as the main sources of error for asynchronous calibration, followed by table height and reconstruction kernel. Based on these results, a correction method was proposed to improve the calibration accuracy.Keywords
Funding Information
- Sinergia Grant from the Swiss National Science Foundation (SNFS-183584)
This publication has 25 references indexed in Scilit:
- Phantomless calibration of CT scans for measurement of BMD and bone strength—Inter-operator reanalysis precisionBone, 2017
- Variation in CT Number and Image Noise Uniformity According to Patient Positioning in MDCTAmerican Journal of Roentgenology, 2017
- Asynchronously Calibrated Quantitative Bone DensitometryJournal of Clinical Densitometry, 2017
- Advanced Virtual Monoenergetic Computed Tomography of Hyperattenuating and Hypoattenuating Liver LesionsInvestigative Radiology, 2015
- The Effect of Quantitative Computed Tomography Acquisition Protocols on Bone Mineral Density EstimationJournal of Biomechanical Engineering, 2015
- Clinical Use of Quantitative Computed Tomography–Based Advanced Techniques in the Management of Osteoporosis in Adults: the 2015 ISCD Official Positions—Part IIIJournal of Clinical Densitometry, 2015
- Novel anthropomorphic hip phantom corrects systemic interscanner differences in proximal femoral vBMDPhysics in Medicine & Biology, 2014
- Simulated increases in body fat and errors in bone mineral density measurements by DXA and QCTJournal of Bone and Mineral Research, 2011
- Helical Multidetector Row Quantitative Computed Tomography (QCT) PrecisionAcademic Radiology, 2009
- The influence of patient centering on CT dose and image noiseMedical Physics, 2007