Reduced Patient Radiation Exposure during Neurodiagnostic and Interventional X-Ray Angiography with a New Imaging Platform
Open Access
- 19 January 2017
- journal article
- research article
- Published by American Society of Neuroradiology (ASNR) in American Journal of Neuroradiology
- Vol. 38 (3) , 442-449
Abstract
BACKGROUND AND PURPOSE: Advancements in medical device and imaging technology as well as accruing clinical evidence have accelerated the growth of the endovascular treatment of cerebrovascular diseases. However, the augmented role of these procedures raises concerns about the radiation dose to patients and operators. We evaluated patient doses from an x-ray imaging platform with radiation dose–reduction technology, which combined image noise reduction, motion correction, and contrast-dependent temporal averaging with optimized x-ray exposure settings. MATERIALS AND METHODS: In this single-center, retrospective study, cumulative dose-area product inclusive of fluoroscopy, angiography, and 3D acquisitions for all neurovascular procedures performed during a 2-year period on the dose-reduction platform were compared with a reference platform. Key study features were the following: The neurointerventional radiologist could select the targeted dose reduction for each patient with the dose-reduction platform, and the statistical analyses included patient characteristics and the neurointerventional radiologist as covariates. The analyzed outcome measures were cumulative dose (kerma)-area product, fluoroscopy duration, and administered contrast volume. RESULTS: A total of 1238 neurointerventional cases were included, of which 914 and 324 were performed on the reference and dose-reduction platforms, respectively. Over all diagnostic and neurointerventional procedures, the cumulative dose-area product was significantly reduced by 53.2% (mean reduction, 160.3 Gy × cm2; P < .0001), fluoroscopy duration was marginally significantly increased (mean increase, 5.2 minutes; P = .0491), and contrast volume was nonsignificantly increased (mean increase, 15.3 mL; P = .1616) with the dose-reduction platform. CONCLUSIONS: A significant reduction in patient radiation dose is achievable during neurovascular procedures by using dose-reduction technology with a minimal impact on workflow.Keywords
This publication has 28 references indexed in Scilit:
- Significant Radiation Dose Reduction in the Hybrid Operating Room Using a Novel X-ray Imaging TechnologyEuropean Journal of Vascular and Endovascular Surgery, 2015
- Evaluation of a Noise Reduction Imaging Technology in Iliac Digital Subtraction Angiography: Noninferior Clinical Image Quality with Lower Patient and Scatter DoseJournal of Vascular and Interventional Radiology, 2015
- Patient radiation dose reduction using an X-ray imaging noise reduction technology for cardiac angiography and interventionHeart and Vessels, 2015
- Image Noise Reduction Algorithm for Digital Subtraction Angiography: Clinical ResultsRadiology, 2013
- Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patientsNeuroradiology, 2013
- Radiation Exposure in Vascular Angiographic ProceduresJournal of Vascular and Interventional Radiology, 2012
- Staff Doses in Interventional Radiology: A National SurveyJournal of Vascular and Interventional Radiology, 2012
- Neurointerventional Procedural Volume per Hospital in United StatesStroke, 2012
- Guidelines for Patient Radiation Dose ManagementJournal of Vascular and Interventional Radiology, 2009
- International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusionThe Lancet, 2005