Abstract
Background To develop a novel custom-made navigational template for accurate prosthesis implantation in total hip resurfacing arthroplasty (THRA) by computer-aided technology. Methods The template was produced based on data preoperatively acquired with computed tomography (CT) scan. The position of the drill guide was chosen as the anatomical axis of the femoral neck which was defined by the ligature between the center of the femoral head and another point on the centerline of proximal femoral neck. And the final direction of the drill guide was confirmed by a small valgus angle. The surface of the template was constructed as the inverse of the femoral neck surface. Then each template was formed from acrylate resin by using rapid prototyping (RP) technique. Finally, the template was tested in 18 cadavers scheduled for THRA and postoperative medical imaging was used to evaluate the accuracy and validity of the template. Results The template had a near-prefect fit with the femoral neck surface. There were no guide failures. Postoperative evaluation revealed that the Kirschner wires pass through the center of the femoral head, have a relative valgus angle to the central axis of the femoral neck and have no obvious varus-valgus angles or anterior/posterior inclination angles. So, the Kirschner wires achieved the expected implant requirement. Conclusion This study shows a novel way of guiding the prosthesis implantation accurately in THRA. The template is an excellent medium for preoperative planning and intraoperative technique, and embodies the development tendency of Computer Assisted Surgery.