Abstract
This paper reports on the development, characterization, and validation of neural probes serving the growing needs of neuroscience for miniaturized tools enabling simultaneous high-resolution recording of neural activity in multiple brain areas. The probes consist of a needle-shaped shaft with a cross-section of 100x 50,μm² and a length of 10 or 5 mm emerging from a base with dimensions of only 0.55x 1.8 mm². The shafts carry 1600 and 800 recording sites, respectively, grouped into 50 (respectively 25) blocks of 4x 8 electrodes with an area of 17x 17,μm² each, a pitch of 20 μm, and an electrode-to-electrode spacing of 3 μm. The probes are fabricated using a commercial 0.18 μm CMOS process followed by dedicated metallization, passivation, and microfabrication steps. Neural signals are accessible through 32 analog output channels via a hierarchical digital addressing scheme implementing an advanced electronic depth control concept giving the option of multiple scanning modes and offering a switching time of 416 μs at a clock frequency of 1 MHz. All output channels are shielded against each other, whereby crosstalk between neighboring channels is measured to be -58 dB at 1 kHz. Absolute impedance values at 1 kHz of single IrOₓ and Pt electrodes are 230± 38 kΩ and 2.2± 0.3 MΩ, respectively. In vivo recordings taking advantage of the new addressing concept for high-resolution recordings from multiple brain regions were successfully performed in anesthetized rats. [2018-0152]
Funding Information
  • Cluster of Excellence BrainLinks-BrainTools through the German Research Foundation (EXC 1086)

This publication has 40 references indexed in Scilit: