Simple method for MR gradient system characterization and k‐space trajectory estimation
Open Access
- 21 December 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 68 (1) , 120-129
Abstract
Fast imaging trajectories are used in MRI to speed up the acquisition process, but imperfections in the gradient system create artifacts in the reconstructed images. Artifacts result from the deviation between k‐space trajectories achieved on the scanner and their original prescription. Measuring or approximating actual k‐space trajectories with predetermined gradient timing delays reduces the artifacts, but are generally based on a specific trajectory and scan orientation. A single linear time‐invariant characterization of the gradient system provides a method to predict k‐space trajectories scanned in arbitrary orientations through convolution. This is done efficiently, by comparing the Fourier transforms of the input and measured waveforms of a single high‐bandwidth test gradient waveform. This new method is tested for spiral, interleaved echo‐planar, and three‐dimensional cones imaging, demonstrating its ability to reduce reconstructed image artifacts for various k‐space trajectories. Magn Reson Med, 2012.Keywords
This publication has 24 references indexed in Scilit:
- Fast concomitant gradient field and field inhomogeneity correction for spiral cardiac imagingMagnetic Resonance in Medicine, 2011
- Fast, simple gradient delay estimation for spiral MRIMagnetic Resonance in Medicine, 2010
- Characterizing and correcting gradient errors in non‐cartesian imaging: Are gradient errors linear time‐invariant (LTI)?Magnetic Resonance in Medicine, 2009
- Estimation of k‐space trajectories in spiral MRIMagnetic Resonance in Medicine, 2009
- Characterization and correction of system delays and eddy currents for MR imaging with ultrashort echo‐time and time‐varying gradientsMagnetic Resonance in Medicine, 2009
- Convolution kernel design and efficient algorithm for sampling density correctionMagnetic Resonance in Medicine, 2009
- Improved k‐space trajectory measurement with signal shiftingMagnetic Resonance in Medicine, 2007
- Discrete time techniques for time delay estimationIEEE Transactions on Signal Processing, 1993
- Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systemsJournal of Magnetic Resonance (1969), 1990
- Optimization of eddy-current compensationJournal of Magnetic Resonance (1969), 1990