Evaluation of hemolysis index thresholds for 18 biochemistry assays: implications for laboratory-developed tests in the era of the IVDR
Open Access
- 27 May 2025
- journal article
- research article
- Published by Walter de Gruyter GmbH in Journal of Laboratory Medicine
Abstract
Objectives: Hemolysis in laboratory samples is a frequent error that may lead to clinical misinterpretation and incorrect patient treatment. Thus, modern laboratory analyzers are able to assess the hemolysis index (HI). To prevent reporting of erroneous results, manufacturers are obliged to provide reliable HI threshold values. The aim of this study was to proof the quality of manufacturer’s given HI thresholds. Methods: Dilution series with defined degrees of hemolysis were prepared by using blood samples from voluntary participants (n=77). In addition to the HI in each dilution, 18 measurands, namely sodium, potassium, chloride, aspartate aminotransferase (ASAT), alanine aminotransferase (ALAT), gamma-glutamyl transferase (GGT), lactate dehydrogenase (LDH), haptoglobin, total bilirubin, direct bilirubin, glutamate dehydrogenase (GLDH), creatine kinase (CK), iron, amylase, phosphate, total protein, enzymatic creatinine and high-sensitive troponin T, were measured on a cobas®pro analyzer. Thresholds for maximum deviation of measurement results of 5 , 10 and 20 % were calculated. Then we determined cut-offs as the 5th percentile and median. Results: Using potassium as an example we found HI thresholds at 57 (5th percentile) and 70 (median) with a corresponding result deviation of 5 % while the manufacturer HI threshold is given at 20. Thus, a higher HI threshold might be tolerated. Conclusions: We established HI thresholds for 18 biochemistry assays. Eight assays showed considerable deviations of which six may have potential clinical relevance like potassium and high-sensitive troponin T. Optimizing thresholds can help to reduce the risk of unnecessary blocking of results and preventing considerably impaired results from being released.Keywords
This publication has 17 references indexed in Scilit:
- Plasma hemoglobin: A method comparison of six assays for hemoglobin and hemolysis index measurementInternational Journal of Laboratory Hematology, 2021
- Automated Measurement of Plasma Cell-Free Hemoglobin Using the Hemolysis Index Check FunctionThe Journal of Applied Laboratory Medicine, 2020
- Determination of hemolysis index thresholds for biochemical tests on Siemens Advia 2400 chemistry analyzerJournal of Clinical Laboratory Analysis, 2019
- 2017 ESC Guidelines for the management of acute myocardial infarction in patients presenting with ST-segment elevationEuropean Heart Journal, 2017
- Validation of Hemolysis Index Thresholds Optimizes Detection of Clinically Significant HemolysisAmerican Journal of Clinical Pathology, 2015
- Can the Roche hemolysis index be used for automated determination of cell-free hemoglobin? A comparison to photometric assaysClinical Biochemistry, 2013
- Time-Dependent Degradation Pattern of Cardiac Troponin T Following Myocardial InfarctionClinical Chemistry, 2013
- Samples: From the Patient to the Laboratory2003
- Hemolyzed Specimens: A Reason for Rejection or a Clinical Challenge?Clinical Chemistry, 2000
- Graphical comparisons of interferences in clinical chemistry instrumentation.Clinical Chemistry, 1986