Abstract
Feature Article Tomokazu Fukui, MD; Shigeo Fukunishi, MD; Shoji Nishio, MD; Nao Shibanuma, MD; Shinichi Yoshiya, MD We have been using an image-free total hip arthroplasty (THA) navigation system (OrthoPilot; Aesculap, Tuttlingen, Germany) to ensure accurate and reproducible acetabular cup orientation. In this study, the accuracy of this system in the assessment of acetabular cup orientation was evaluated by comparing the intraoperative inclination and anteversion angles presented by the navigation system and the corresponding postoperative values obtained by computed tomography measurement. In the intraoperative accuracy analysis, we additionally examined the influence of factors such as body mass index and soft tissue thickness on assessment error. Intraoperative and postoperative results obtained from 115 consecutive navigated THAs were compared and analyzed. In both inclination and anteversion angles, good agreement was observed; a discrepancy of more than 5° was observed in 1 and 3 cases, respectively. In the analysis of factors potentially affecting the accuracy of the intraoperative assessment, no correlation between each parameter and the intraoperative and postoperative discrepancy was demonstrated. Drs Fukui, Fukunishi, Nishio, and Yoshiya are from the Department of Orthopaedic Surgery, Hyogo College of Medicine, Nishinomiya, and Dr Shibanuma is from the Department of Orthopaedic Surgery, Kobe Kaisei Hospital, Kobe, Japan. Drs Fukui, Fukunishi, Nishio, Shibanuma, and Yoshiya have no relevant financial relationships to disclose. Correspondence should be addressed to: Tomokazu Fukui, MD, Department of Orthopaedic Surgery, Hyogo College of Medicine, 1-1 Mukogawa-cho Nishinomiya, Hyogo 663-8501, Japan. Use of a computer-assisted navigation system in total hip arthroplasty (THA) has been introduced with expectations of achieving optimum implant alignment.1–8 Current THA navigation systems can be divided into 2 groups based on the operation principle: image-based or image-free. In the image-free system, the position and orientation of the patient’s pelvis are recognized intraoperatively by palpation and registration of the bony landmarks. Thus, a discrepancy between the locations of the bony landmark and the registered point may cause assessment errors. In several clinical studies, the accuracy of the image-free THA navigation system has been assessed and reported.3–5,9–12 In most studies, satisfactory agreement between the intraoperative navigation results and the postoperative values obtained from radiologic examination has been reported.3–5,13,14 We have been using the OrthoPilot image-free navigation system (B. Braun Aesculap, Tuttlingen, Germany) since 2006. In order to examine the accuracy and potential source of error in the use of this navigation system, we conducted a study comparing the intraoperative cup positioning angles shown by the navigation and the corresponding values calculated from postoperative computed tomography (CT) data. In the CT evaluation, we analyzed Digital Imaging and Communications in Medicine (DICOM) data using the Medical Image Processing, Analysis and Visualization (MIPAV) computer program developed by the National Institutes of Health (Bethesda, Maryland), which allows the consistent creation of an analysis coordinate system.15,16 We reported preliminary results of the analysis for the initial 25 THAs in 2008.17 Although a reasonably high accuracy with the use of this system was observed, the small sample size in the study precluded us from analyzing the factors affecting the accuracy and presenting a conclusive statement. One of the potential factors leading to assessment error in the use of the image-free system is inaccurate localization of the bony landmarks.12,18,19 In particular, the presence of thick subcutaneous tissue in an obese patient can cause a discrepancy. In this study, we increased the number of patients and examined the effects of soft tissue on assessment error in the intraoperative evaluation. Moreover, it was expected that the accuracy of this system could be more substantially assessed than in our previous report17 due to the analysis of a larger patient population. In 2005, we began using the OrthoPilot CT-free navigation system. After an initial series of 38 hips, we designed and started this study in April 2006 after approval by the Institutional Review Board and with informed consent from the patients. From April 2006 to March 2008, both acetabular alignment data based on navigation system and acetabular alignment data based on postoperative CT evaluation were obtained and compared in 115 consecutive navigated THAs in 110 patients. Mean patient age was 63.4 years (range, 21–88 years). Fifteen patients were men and 95 were women. Preoperative diagnosis was osteoarthritis in 85, osteonecrosis in 25, rheumatoid arthritis in 3, and hemophilic arthropathy in 2 patients. Preoperatively, 2 metal markers were attached to the bilateral anterosuperior iliac spines, and the location of the markers was confirmed by a plain anteroposterior pelvic radiograph (Figure 1). These markers showed the location of the bony landmarks both in the intraoperative and postoperative evaluations. Before surgery, the tracker was inserted and fixed to the iliac crest through a stab incision with the patient in the supine position. As the anatomic landmarks, the bilateral anterosuperior iliac spines and the upper margin of the symphysis were located by percutaneous palpation of the markers and bony prominence. The reference plane in the analysis was...

This publication has 18 references indexed in Scilit: