Background An ablation catheter capable of contact force (CF) and local impedance (LI) monitoring (IntellaNav StablePoint, Boston Scientific) has been recently launched. We evaluated the relationship between the CF and LI values during radiofrequency catheter ablation (RFCA) along the cavotricuspid isthmus (CTI). Methods Fifty consecutive subjects who underwent a CTI-RFCA using IntellaNav StablePoint catheters were retrospectively studied. The initial CF and LI at the start of the RF applications and mean CF and minimum LI during the RF applications were measured. The absolute and percentage LI drops were calculated as the difference between the initial and minimum LIs and 100 × absolute LI drop/initial LI, respectively. Results We analyzed 602 first-pass RF applications. A weak correlation was observed between the initial CF and LI (r = 0.13) and between the mean CF and LI drops (r = 0.22). The initial LI and absolute and percentage LI drops were greater at effective ablation sites than ineffective ablation sites (median, 151 vs. 138 Ω, 22 vs. 14 Ω, and 14.4% vs. 9.9%; p < .001), but the initial and mean CF did not differ. At optimal cutoffs of 21 Ω and 10.8% for the absolute and percentage LI drops according to the receiver-operating characteristic analysis, the sensitivity, and specificity for predicting an effective ablation were 57.4% and 88.9% and 80.0%, and 61.1%, respectively. Conclusions The effective sites during the CF-guided CTI-RFCA had greater initial LI and LI drops than the ineffective sites. Absolute and percentage LI drops of 21 Ω and 10.8% may be appropriate targets for an effective ablation.Keywords
This publication has 23 references indexed in Scilit:
- Use of impedance-based catheter tip-to-tissue contact assessment (electroanatomic coupling index, ECI) in typical right atrial flutter ablationJournal of Interventional Cardiac Electrophysiology, 2018
- Novel Measure of Local Impedance Predicts Catheter–Tissue Contact and Lesion FormationCirculation: Arrhythmia and Electrophysiology, 2018
- Characteristics of Cavotricuspid Isthmus Ablation for Atrial Flutter Guided by Novel Parameters Using a Contact Force CatheterJournal of Cardiovascular Electrophysiology, 2016
- Contact-force-guided vs. contact-force-blinded catheter ablation of typical atrial flutter: a prospective studyEP Europace, 2016
- Use of contact force sensing technology during radiofrequency ablation reduces recurrence of atrial fibrillation: A systematic review and meta-analysisHeart Rhythm, 2015
- Meta-Analysis of Ablation of Atrial Flutter and Supraventricular TachycardiaThe American Journal of Cardiology, 2009
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- Double potentials along the ablation line as a guide to radiofrequency ablation of typical atrial flutterJACC, 2001
- Differential Pacing for Distinguishing Block From Persistent Conduction Through an Ablation LineCirculation, 2000
- Electrophysiological Effects of Catheter Ablation of Inferior Vena Cava–Tricuspid Annulus Isthmus in Common Atrial FlutterCirculation, 1996