Aims: Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear.Methods: A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay.Results: At mean six years’ follow-up, the PE wear rate was 0.08 mm/year (95% confidence interval 0.06 to 0.09 mm/year). PE inlay external rotation was below the precision limit and did not influence PE wear. Varus knee alignment did not influence PE wear (p = 0.874), but increased tibial component total translation (p = 0.041).Conclusion: The PE inlay was well fixed and there was no relationship between PE stability and PE wear. The PE wear rate was low and similar in the medial and lateral compartments. Varus knee alignment did not influence PE wear.Cite this article: Bone Joint Res 2024;13(5):226–236.Keywords
This publication has 28 references indexed in Scilit:
- Editorial: What is Varus or Valgus Knee Alignment?Clinical Orthopaedics and Related Research, 2010
- A new type of model-based Roentgen stereophotogrammetric analysis for solving the occluded marker problemJournal of Biomechanics, 2005
- Backside Wear of Polyethylene Tibial InsertsJournal of Bone and Joint Surgery, 2005
- Guidelines for standardization of radiostereometry (RSA) of implantsActa Orthopaedica, 2005
- Backside Wear of Miller-Galante I and Insall-Burstein II Tibial InsertsClinical Orthopaedics and Related Research, 2004
- Assessment of Backside Wear From the Analysis of 55 Retrieved Tibial InsertsClinical Orthopaedics and Related Research, 2002
- Wear of high-dose gamma-irradiated polyethylene in total hip replacementsJournal of Orthopaedic Science, 2000
- The reliability of the American Knee Society ScoreActa Orthopaedica, 2000
- Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses.1995
- Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prosthesesJournal of Biomedical Materials Research, 1983