Abstract
The aim of this study was to evaluate the feasibility of an NMR-on-a-chip sensor as local field probe for contactless monitoring of respiratory and cardiac motion-induced field changes and to investigate their applicability for motion synchronization in MRI. Cardiac and respiratory motion-induced field changes were recorded at 7 positions around the torso of 6 volunteers. The data were analysed regarding cardiac and respiratory components and the most promising location for assessment of respiratory motion identified. Further, the sensor-derived signal was correlated with the vendor-provided respiratory belt. The cranial parasternal positions were identified as excellent candidates for mapping cardiac motion, whereas the anterior locations at height of the diaphragm appear most promising for mapping of the respiration. Realtime analysis of the recorded signals was successfully applied for respiratory synchronization of the MR system and showed very good correlation with the respiratory belt with less fluctuation and shorter transient phase during breath-holding. NMR-on-a-chip sensors are sensitive enough for monitoring physiological motion and can be applied for respiratory motion synchronization. Depending on the positioning of the sensor, cardiac as well as respiratory induced field changes can be contactless mapped and used for synchronization of the MRI.