A direct comparison of multi-energy x-ray and proton CT for imaging and relative stopping power estimation of plastic and ex-vivo phantoms
Open Access
- 30 August 2024
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 69 (17) , 175021
Abstract
Objective. Proton therapy administers a highly conformal dose to the tumour region, necessitating accurate prediction of the patient’s 3D map of proton relative stopping power (RSP) compared to water. This remains challenging due to inaccuracies inherent in single-energy computed tomography (SECT) calibration. Recent advancements in spectral x-ray CT (xCT) and proton CT (pCT) have shown improved RSP estimation compared to traditional SECT methods. This study aims to provide the first comparison of the imaging and RSP estimation performance among dual-energy CT (DECT) and photon-counting CT (PCCT) scanners, and a pCT system prototype. Approach. Two phantoms were scanned with the three systems for their performance characterisation: a plastic phantom, filled with water and containing four plastic inserts and a wood insert, and a heterogeneous biological phantom, containing a formalin-stabilised bovine specimen. RSP maps were generated by converting CT numbers to RSP using a calibration based on low- and high-energy xCT images, while pCT utilised a distance-driven filtered back projection algorithm for RSP reconstruction. Spatial resolution, noise, and RSP accuracy were compared across the resulting images. Main results. All three systems exhibited similar spatial resolution of around 0.54 lp/mm for the plastic phantom. The PCCT images were less noisy than the DECT images at the same dose level. The lowest mean absolute percentage error (MAPE) of RSP, ( 0.28 ± 0.07 ) % , was obtained with the pCT system, compared to MAPE values of ( 0.51 ± 0.08 ) % and ( 0.80 ± 0.08 ) % for the DECT- and PCCT-based methods, respectively. For the biological phantom, the xCT-based methods resulted in higher RSP values in most of the voxels compared to pCT. Significance. The pCT system yielded the most accurate estimation of RSP values for the plastic materials, and was thus used to benchmark the xCT calibration performance on the biological phantom. This study underlined the potential benefits and constraints of utilising such a novel ex-vivo phantom for inter-centre surveys in future.Keywords
This publication has 47 references indexed in Scilit:
- Comparative accuracy and resolution assessment of two prototype proton computed tomography scannersMedical Physics, 2022
- A comparison of proton stopping power measured with proton CT and x‐ray CT in fresh postmortem porcine structuresMedical Physics, 2021
- Experimental comparison of photon versus particle computed tomography to predict tissue relative stopping powersMedical Physics, 2021
- Technical Note: A fast and monolithic prototype clinical proton radiography system optimized for pencil beam scanningMedical Physics, 2021
- Technical Note: CT calibration for proton treatment planning by cross‐calibration with proton CT dataMedical Physics, 2020
- Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233Medical Physics, 2019
- Theoretical and experimental analysis of photon counting detector CT for proton stopping power predictionMedical Physics, 2018
- Material elemental decomposition in dual and multi‐energy CT via a sparsity‐dictionary approach for proton stopping power ratio calculationMedical Physics, 2018
- Simplified derivation of stopping power ratio in the human body from dual‐energy CT dataMedical Physics, 2017
- A simple formulation for deriving effective atomic numbers via electron density calibration from dual-energy CT data in the human bodyMedical Physics, 2017