A 4-Channel Neural Stimulation IC Design With Charge Balancing and Multiple Current Output Modes
- 22 September 2023
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. 17 (5) , 1037-1049
Abstract
This paper proposes a neural stimulation integrated circuit design with multiple current output modes. In the cathodic stimulation phase and anodic stimulation phase, each output current waveform can be independently selected to either exponential waveform or square wave, so the stimulator holds four stimulation modes. To minimize the headroom voltage of the output stage and enhance the power efficiency of the proposed stimulator, we introduce the exponentially decaying current which is realized by the exponential current generation circuit in this work. It can enhance the longer duration of the stimulation pulse as well. In case the residual charge may cause harm to patients, a charge balancing technique is implemented in this work for all operation modes. The four-channel stimulator IC is implemented in a 180-nm CMOS process, occupying a core area of 1.93 mm 2 . The measurement results show that the proposed stimulator realized a maximum power efficiency of 91.3% and the maximum stimulation duration is 3 times larger than previous works. Moreover, even in exponential output waveform mode, the maximum residual charge in a single cycle is only 255 pC due to the proposed charge balancing technique. The experiment results based on the PBS solution also show that the stimulator IC can remove residual charges within 60 μs, and the electrode voltage remains stable within a safe range under multicycle stimulation.Keywords
Funding Information
- Beijing Natural Science Foundation Program (4222062)
- National Natural Science Foundation of China (61806012)
This publication has 34 references indexed in Scilit:
- A 16-channel 24-V 1.8-mA power efficiency enhanced neural/muscular stimulator with exponentially decaying stimulation current2015
- An Energy-Efficient, Adiabatic Electrode Stimulator With Inductive Energy Recycling and Feedback Current RegulationIEEE Transactions on Biomedical Circuits and Systems, 2011
- A Stimulator ASIC Featuring Versatile Management for Vestibular ProsthesesIEEE Transactions on Biomedical Circuits and Systems, 2011
- A Wireless Integrated Circuit for 100-Channel Charge-Balanced Neural StimulationIEEE Transactions on Biomedical Circuits and Systems, 2009
- Responsive Cortical Stimulation for the Treatment of EpilepsyNeurotherapeutics, 2008
- Tissue Damage by Pulsed Electrical StimulationIEEE Transactions on Biomedical Engineering, 2007
- A Low-Power Blocking-Capacitor-Free Charge-Balanced Electrode-Stimulator Chip With Less Than 6 nA DC Error for 1-mA Full-Scale StimulationIEEE Transactions on Biomedical Circuits and Systems, 2007
- Electrical stimulation of excitable tissue: design of efficacious and safe protocolsJournal of Neuroscience Methods, 2005
- Deep Brain Stimulation for Psychiatric DisordersJournal Of Clinical Neurophysiology, 2004
- An externally powered, multichannel, implantable stimulator-telemeter for control of paralyzed muscleIEEE Transactions on Biomedical Engineering, 1998