Detection of patient movement during CBCT examination using video observation compared with an accelerometer-gyroscope tracking system
- 1 February 2017
- journal article
- research article
- Published by Oxford University Press (OUP) in Dentomaxillofacial Radiology
- Vol. 46 (2) , 20160289
Abstract
To compare video observation (VO) with a novel three-dimensional registration method, based on an accelerometer-gyroscope (AG) system, to detect patient movement during CBCT examination. The movements were further analyzed according to complexity and patient age. In 181 patients (118 females/63 males; age average 30 years, range: 9–84 years), 206 CBCT examinations were performed, which were video-recorded during examination. An AG was, at the same time, attached to the patient head to track head position in three dimensions. Three observers scored patient movement (yes/no) by VO. AG provided movement data on the x-, y- and z-axes. Thresholds for AG-based registration were defined at 0.5, 1, 2, 3 and 4 mm (movement distance). Movement detected by VO was compared with that registered by AG, according to movement complexity (uniplanar vs multiplanar, as defined by AG) and patient age (≤15, 16–30 and ≥31 years). According to AG, movement ≥0.5 mm was present in 160 (77.7%) examinations. According to VO, movement was present in 46 (22.3%) examinations. One VO-detected movement was not registered by AG. Overall, VO did not detect 71.9% of the movements registered by AG at the 0.5-mm threshold. At a movement distance ≥4 mm, 20% of the AG-registered movements were not detected by VO. Multiplanar movements such as lateral head rotation (72.1%) and nodding/swallowing (52.6%) were more often detected by VO in comparison with uniplanar movements, such as head lifting (33.6%) and anteroposterior translation (35.6%), at the 0.5-mm threshold. The prevalence of patients who move was highest in patients younger than 16 years (64.3% for VO and 92.3% for AG-based registration at the 0.5-mm threshold). AG-based movement registration resulted in a higher prevalence of patient movement during CBCT examination than VO-based registration. Also, AG-registered multiplanar movements were more frequently detected by VO than uniplanar movements. The prevalence of patients who move was highest in patients younger than 16 years.Keywords
This publication has 17 references indexed in Scilit:
- Movement characteristics in young patients and the impact on CBCT image qualityDentomaxillofacial Radiology, 2016
- Factors affecting patient movement and re-exposure in cone beam computed tomography examinationOral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 2015
- Radiographic observers' ability to recognize patient movement during cone beam CTDentomaxillofacial Radiology, 2014
- Can Dental Cone Beam Computed Tomography Assess Bone Mineral Density?Journal of Bone Metabolism, 2014
- Movement of the patient and the cone beam computed tomography scanner: objectives and possible solutionsOral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 2013
- Dental cone beam CT image quality possibly reduced by patient movementDentomaxillofacial Radiology, 2013
- Dose distribution for dental cone beam CT and its implication for defining a dose indexDentomaxillofacial Radiology, 2012
- Effective dose range for dental cone beam computed tomography scannersEuropean Journal of Radiology, 2012
- Artefacts in CBCT: a reviewDentomaxillofacial Radiology, 2011
- Reduction of the negative influence of patient motion on quality of CBCT scanMedical Hypotheses, 2010