Dark-Adapted Pupil Diameter as a Function of Age Measured with the NeurOptics Pupillometer
Authors:
- Karl C. BentleyKBKarl C. Bentley
- Aleem I. MughalAMAleem I. Mughal
- Hari BodhireddyHBHari Bodhireddy
- Sandra M. BrownSBSandra M. Brown
- 1 March 2011
- journal article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 27 (3) , 202-207
Abstract
Purpose: To measure the dark-adapted pupil diameter of normal research participants in their second through ninth decades of life using the NeurOptics pupillometer (Neuroptics Inc). Methods: Individuals aged 18 to 80 years with no history of eye disease or injury, intraocular surgery, or use of systemic antihistamines or opiates were recruited. After 2 minutes of adaptation at 1 lux illumination, the right dark-adapted pupil diameter was measured using the NeurOptics pupillometer, with accommodation controlled by distance fixation. The NeurOptics pupillometer reported a mean dark-adapted pupil diameter and a standard deviation of the mean, which were analyzed as a function of age-decade. Results: Two-hundred sixty-three individuals participated. For participants aged 18 to 19 years (n=6), the mean dark-adapted pupil diameter was 6.85 mm (range: 5.6 to 7.5 mm); 20 to 29 years (n=66), 7.33 mm (range: 5.7 to 8.8 mm); 30 to 39 years (n=50), 6.64 mm (range: 5.3 to 8.7 mm); 40 to 49 years (n=51), 6.15 mm (range: 4.5 to 8.2 mm); 50 to 59 years (n=50), 5.77 mm (range: 4.4 to 7.2 mm); 60 to 69 years (n=30), 5.58 mm (range: 3.5 to 7.5 mm); 70 to 79 years (n=6), 5.17 mm (range: 4.6 to 6.0 mm); and 80 years (n=4), 4.85 mm (range: 4.1 to 5.3 mm). These values were consistent with studies using infrared photography. The standard deviation was >0.1 mm in 10 (3.8%) participants, all of whom were younger than 55 years. Conclusions: The dark-adapted pupil diameter is an important clinical variable when planning refractive surgery. Surgeons can compare a patient’s dark-adapted pupil diameter with the results of this population study to identify outlier measurements, which may be erroneous, and repeat testing prior to surgery.Keywords
This publication has 32 references indexed in Scilit:
- Clinical performance of a handheld digital infrared monocular pupillometer for measurement of the dark-adapted pupil diameterJournal of Cataract & Refractive Surgery, 2010
- Comparison of 2 multiple-measurement infrared pupillometers to determine scotopic pupil diameterJournal of Cataract & Refractive Surgery, 2006
- Pupil size and corneal laser surgeryCurrent Opinion in Ophthalmology, 2006
- Assessment of pupil size under different light intensities using the Procyon pupillometerJournal of Cataract & Refractive Surgery, 2006
- Measuring room illuminanceJournal of Cataract & Refractive Surgery, 2006
- Comparison of Colvard pupillometer and infrared digital photography for measurement of the dark-adapted pupil diameterJournal of Cataract & Refractive Surgery, 2005
- Standardization is required to evaluate optical quality after LASIKJournal of Cataract & Refractive Surgery, 2005
- Risk factors for night vision complaints after LASIK for myopiaOphthalmology, 2004
- Night vision after LASIK: the pupil proclaims innocenceOphthalmology, 2004
- Dependence on Age of the Diameter of the Pupil in the DarkNature, 1958