Abstract
Background Young implantable cardioverter-defibrillator (ICD) patients are prone to complications and inappropriate shocks (IAS). The subcutaneous ICD (S-ICD) may avoid lead-related complications. This study aims to describe the incidence and nature of device-related complications in young transvenous ICD (TV-ICD) and S‑ICD patients. Methods Single-chamber TV-ICD and S‑ICD patients up to and including the age of 25 years implanted between 2002 and 2015 were retrospectively analysed. Complications were defined as device-related complications requiring surgical intervention. IAS were defined as shocks for anything other than ventricular tachycardia or ventricular fibrillation. Follow-up data were collected 5 years post-implantation. Kaplan-Meier estimates for complications at 5‑year follow-up were calculated with a corresponding 95% confidence interval. Results Eighty-one patients (46 TV-ICD, 35 S-ICD) were included (median age 19.0 (IQR 16.0–23.0) and 16.5 (IQR 13.0–20.2) years respectively). Median follow-up was 60 and 40 months respectively. All-cause complication rate was 34% in the TV-ICD group and 25% in the S‑ICD group (p = 0.64). TV-ICD patients had more lead complications: 23% (10–36%) versus 0% (p = 0.02). The rate of infections did not differ between TV-ICD and S‑ICD: 2% (0–6%) versus 10% (0–21%) (p = 0.15). No systemic infections occurred in the S‑ICD patients. The rates of IAS were similar, TV-ICD 22% (9–35%) versus S‑ICD 14% (0–30%) (p = 0.40), as were those for appropriate shocks: 25% (11–39%) versus 27% (6–48%) (p = 0.92). Conclusion The rates of all-cause complications in this cohort were equal, though the nature of the complications differed. S‑ICD patients did not suffer lead failures or systemic infections. An era effect is present between the two groups.Keywords
This publication has 17 references indexed in Scilit:
- Implantable cardioverter-defibrillator harm in young patients with inherited arrhythmia syndromes: A systematic review and meta-analysis of inappropriate shocks and complicationsHeart Rhythm, 2016
- 2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac deathEuropean Heart Journal, 2015
- Safety and Efficacy of the Totally Subcutaneous Implantable DefibrillatorJACC, 2015
- Clinical Experience of Subcutaneous and Transvenous Implantable Cardioverter Defibrillators in Children and TeenagersPacing and Clinical Electrophysiology, 2013
- Multi-Institutional Study of Implantable Defibrillator Lead Performance in Children and Young AdultsCirculation, 2013
- Pediatric ICD utilization in the United States from 1997 to 2006Heart Rhythm, 2011
- Lead Extraction in Pediatric and Congenital Heart Disease PatientsCirculation: Arrhythmia and Electrophysiology, 2010
- Risks and Challenges of Implantable Cardioverter-Defibrillators in Young AdultsProgress in Cardiovascular Diseases, 2008
- Annual Rate of Transvenous Defibrillation Lead Defects in Implantable Cardioverter-Defibrillators Over a Period of >10 YearsCirculation, 2007
- Transvenous Defibrillation Leads: High Incidence of Failure During Long‐Term Follow‐UpJournal of Cardiovascular Electrophysiology, 2003