BACKGROUND: The aim is to quantify the electrode insertion trauma-induced hearing loss (EITHL), its risk factors, and its impact on speech outcomes in prelingually deafened children undergoing cochlear implantation. METHODS: This was a prospective, observational study conducted at a single center between 2021 and 2024. Forty children aged 1–5 years with severe-to-profound sensorineural hearing loss underwent cochlear implantation using either the cochleostomy or round-window (RW) techniques. Auditory steady-state response (ASSR) thresholds were utilized to assess hearing preservation levels. Speech outcomes were evaluated by measuring word recognition scores (WRS) at 6, 12, and 24 months. The correlation between electrode insertion depth, hearing preservation, and speech outcomes was analyzed. RESULTS: All participants achieved Grade 1 hearing preservation (>75%). The mean hearing preservation rates at 3, 6, 9, and 12 months were 93.2%, 92.3%, 92.9%, and 92.1%, respectively. Although the RWT demonstrated better hearing preservation than the CS technique, the difference was not statistically significant. A significant increase in low-frequency hearing thresholds was observed over time, with 12-month values of 84.9 ± 3.5 dB (250 Hz), 90.2 ± 3.7 dB (500 Hz), and 92.4 ± 4.0 dB (1000 Hz). A negative correlation was found between the depth of electrode insertion and hearing preservation (r=−0.45, P=.03). Word recognition scores improved over time, with bilateral implant recipients showing significantly higher scores (P < .00). CONCLUSION: Cochlear implantation via both CS and RW approaches preserves residual hearing in prelingually deafened children, with RW insertion demonstrating superior preservation. Deeper electrode insertion is associated with poorer hearing preservation, while better hearing preservation correlates with improved speech outcomes.Keywords
This publication has 20 references indexed in Scilit:
- Round Window Versus Cochleostomy Technique in Cochlear ImplantationOtology & Neurotology, 2012
- The Role of Hearing Preservation on Electrical Thresholds and Speech Performances in Cochlear ImplantationOtology & Neurotology, 2012
- Implications of Minimizing Trauma During Conventional Cochlear ImplantationOtology & Neurotology, 2011
- Residual Hearing Preservation After Pediatric Cochlear ImplantationOtology & Neurotology, 2010
- A new cochlear implant electrode design for preservation of residual hearing: a temporal bone studyActa Oto-Laryngologica, 2009
- Histopathologic Assessment of Fibrosis and nEw Bone Formation in Implanted Human Temporal Bones Using 3D ReconstructionOtolaryngology–Head and Neck Surgery, 2009
- An evaluation of the preservation of residual hearing with the Nucleus® Contour Advance™ electrodeActa Oto-Laryngologica, 2009
- Preservation of residual hearing following cochlear implantation: comparison between three surgical techniquesThe Journal of Laryngology & Otology, 2007
- Conservation of Residual Acoustic Hearing After Cochlear ImplantationOtology & Neurotology, 2006
- The influence of residual high-frequency hearing on the outcome in congenitally deaf cochlear implant recipients.2000