Effect of radiation protection apron defect on radiation dose during fluoroscopy
Open Access
- 22 July 2026
- journal article
- Published by IOP Publishing in Journal of Radiological Protection
- Vol. 46 (3) , 031519
Abstract
Occupational exposure should be minimised to a level that is as low as reasonably achievable. Radiation protection aprons and thyroid collars are effective in reducing occupational radiation dose as they typically attenuate at least 90% of scattered radiation. However, cracks and holes may compromise the shielding capacity of the apron. The aim of the study was to quantify change in radiation dose behind defects to support the development of rejection criteria for damaged aprons. Air kerma rates were measured with a calibrated solid‐state dosimeter behind a 0.25 mm Pb‐equivalent apron containing seven circular defects (diameter 1.5–20 mm) and one horizontal crack (50 mm) under two clinical fluoroscopy settings and compared with measurements obtained behind an intact apron. The air kerma rate increased by up to 165 µ Gy h −1 behind the damaged apron. Based on average annual fluoroscopy times per operator at Kuopio University Hospital, the use of damaged apron could result in an annual air kerma increase of 9.7 mGy behind the damaged area. Defect size and location affect the shielding performance of protective aprons. Considering defect position in rejection criteria allows more realistic estimation of the potential impact on staff effective dose and helps avoid unnecessary replacement of protective garments. Based on the present measurements, defects smaller than approximately 1 cm did not produce a meaningful increase in air-kerma values under the tested conditions.Keywords
This publication has 13 references indexed in Scilit:
- Comparative Analysis of Effectiveness of Traditional Lead Aprons versus Newer Generation Lead-free Aprons in Radiation ProtectionJournal of Medical Physics, 2025
- 0.35-mm lead-equivalent aprons provide similar protection to interventionalists compared to 0.5-mm lead-equivalent aprons during fluoroscopically guided interventionsJVS-Vascular Insights, 2024
- Reassessing lead protective garment assessment using ICRP 103 tissue weighting factorsJournal of Radiological Protection, 2023
- Comparison of the radiation protection effect of different radiation protection aprons made of different materialsEuropean Journal of Radiology, 2023
- Integrity of personal radiation protective equipment (PRPE): a 4-year longitudinal follow-up studyInsights into Imaging, 2022
- Quality Assurance of Personal Radiation Shield for Kilovoltage Photon: A Multicentre ExperienceRisk Management and Healthcare Policy, 2021
- A simple quality control tool for assessing integrity of lead equivalent apronsIndian Journal of Radiology and Imaging, 2018
- Integrity test of lead apron and its effect on personnel and carerBangabandhu Sheikh Mujib Medical University Journal, 2018
- Protective aprons in imaging departments: manufacturer stated lead equivalence values require validationEuropean Radiology, 2005
- Inspection of Lead ApronsHealth Physics, 2001