Randoms Estimation for Long Axial Field-of-View PET
- 1 March 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 10 (3) , 401-405
Abstract
The high sensitivity of long axial field-of-view (AFOV) PET scanners has enabled studies over a wide range of count rates and count densities. However, these systems have a large axial acceptance angle that necessitates a wide coincidence window to capture the oblique true coincidences. In addition, the measured delays sinogram is sparse and noisy. We studied four methods of randoms estimation on a long AFOV system to assess their impact on accuracy and image noise: measured delays using a delayed coincidence window (RD), 2D Casey averaging of measured delays (RD-smooth), 2D average of measured delays (RD-ave – the current default method on the PennPET Explorer), and estimation of randoms from singles (RS). We looked at cases with varying count densities, randoms fractions, and non-pure positron emitters. A positive bias observed at low randoms counts for the RD and RD-smooth methods was not seen with the RD-ave or RS methods. For all cases, quantitative results with RS agreed to within 2.5% of the RD-ave method, while RD and RD-smooth estimates showed differences of 5-49%, with larger differences in areas of low uptake. The RS method is a practical technique for list-mode data and list-mode reconstruction by reducing the size of stored list events. It also avoids small approximations in the RDave method. For long AFOV systems, estimating randoms from singles is a practical and accurate method.Keywords
Funding Information
- National Institutes of Health (R01-CA113941)
This publication has 15 references indexed in Scilit:
- Performance evaluation of the PennPET explorer with expanded axial coveragePhysics in Medicine & Biology, 2023
- Total Body PET: Why, How, What for?IEEE Transactions on Radiation and Plasma Medical Sciences, 2020
- Total-body imaging: Transforming the role of positron emission tomographyScience Translational Medicine, 2017
- Randoms from singles estimation for long PET scans2011
- Improving the singles rate method for modeling accidental coincidences in high-resolution PETPhysics in Medicine & Biology, 2010
- Implementation and Evaluation of a 3D PET Single Scatter Simulation with TOF Modeling2006
- Iterative image reconstruction using geometrically ordered subsets with list-mode data2005
- Randoms variance reduction in 3D PETPhysics in Medicine & Biology, 1999
- Evaluation of the Karolinska new positron camera system; the Scanditronix PC2048-15BIEEE Transactions on Nuclear Science, 1990
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1986