Radiation Dose to the Thyroid Gland and Breast From Multidetector Computed Tomography of the Cervical Spine
- 1 November 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 33 (6) , 987-990
Abstract
Purpose: This study aimed to assess the radiation dose reduction that could be achieved using an in-line bismuth shielding over the thyroid gland and breast and to determine the effect of a cervical spine collar on thyroid dose reduction and image noise when performing computed tomography of the cervical spine using automatic tube current modulation. Materials and Methods: An anthropomorphic phantom was scanned using a commercially available 64-channel computed tomographic scanner. A standardized trauma cervical spine protocol was used. Scans were obtained with and without a standard cervical spine immobilization collar and with and without bismuth-impregnated thyroid and breast shields. Thermoluminescent dosimeters were placed over the thyroid gland and breasts for each scan. A paired t test was used to determine whether the skin entry dose differed significantly between the shielded and unshielded thyroid and breast and to determine whether placing the thyroid shield over a cervical immobilization collar resulted in a significant dose reduction. Region of interest of pixel values was used to determine image noise. Results: The average measured skin entry dose for the unshielded thyroid gland was 21.9 mGy (95% confidence interval, 18.9-4.7). With a bismuth shield applied directly over the skin, the dose to the thyroid gland was reduced by 22.5% (P < 0.05). With the bismuth shield applied over the cervical spine collar, the dose reduction to the thyroid was 10.4%, which was not statistically significant (P = 0.16) compared with the dose reduction without the cervical collar. Skin entry dose over the breasts was significant, although they were outside the primary scan range. Without bismuth shielding, the skin entry dose was 1.5 mGy, and with bismuth shielding, the dose was significantly reduced by 36.6% (P < 0.01). Image noise increased most when shielding was used with an immobilization collar. Conclusions: There is a significant dose reduction to the thyroid gland and breasts when a bismuth shield is placed on the skin. The dose saving achieved by placing the shield on the cervical collar is approximately halved compared with placement on the skin, and this did not reach statistical significance, and this was accompanied by an increase on image noise. Bismuth shields should not be used in combination with cervical immobilization collars.Keywords
This publication has 16 references indexed in Scilit:
- Radiation Dose Savings for Adult Pulmonary Embolus 64-MDCT Using Bismuth Breast Shields, Lower Peak Kilovoltage, and Automatic Tube Current ModulationAmerican Journal of Roentgenology, 2009
- Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinationsEuropean Radiology, 2008
- Computed Tomography Use in a Tertiary Care University HospitalJournal of the American College of Radiology, 2008
- Pediatric Chest MDCT Using Tube Current Modulation: Effect on Radiation Dose with Breast ShieldingAmerican Journal of Roentgenology, 2008
- Increasing utilization of computed tomography in the pediatric emergency department, 2000–2006Emergency Radiology, 2007
- Energy imparted‐based estimates of the effect of overscanning on adult and pediatric patient effective doses from multi‐slice computed tomographyMedical Physics, 2007
- Overranging in Multisection CT: Quantification and Relative Contribution to Dose—Comparison of Four 16-Section CT ScannersRadiology, 2007
- Radiation Exposure From Diagnostic Imaging in Severely Injured Trauma Patients2007
- Increasing utilization of computed tomography in the adult emergency department, 2000–2005Emergency Radiology, 2006
- Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image qualityEuropean Radiology, 2006