Characterization of Gas Electron Multiplier‐based detector for external beam radiation therapy dosimetry
Authors:
- 13 January 2021
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (4) , 1931-1940
Abstract
Purpose Electronic Portal Imaging Devices (EPIDs) are commonly installed on modern linear accelerators (LINACs) and are convenient for imaging and, potentially, dosimetry. However, owing to their construction with metal and scintillating layers of high atomic number, they exhibit non‐water‐equivalent response and oversensitivity to low‐energy photons. Therefore, EPIDs are not ideal for dosimetry purposes. Additionally, non‐linearities due to the combined use of scintillators and photodiodes have been reported. Here, an EPID which employs a variable gain Gas Electron Multiplier (GEM) and direct detection of electrons is introduced. To investigate its dosimetric performance, measurements characterizing the novel EPID are performed and compared with measurements from ionization chambers and conventional EPIDs. Methods Linearity, dose rate dependence, field size dependence, off‐axis response and transmission response were measured for all available energy settings (6 MV, 10 MV, 6 MV Flattening Filter Free (FFF) and 10 MVFFF) using three different detector gain settings. Additionally, an evaluation of the ghosting and image lag of the panel was completed. Reference ionization chamber measurements were performed for the off‐axis and transmission response and existing data for conventional EPIDs and ionization chambers from equivalent measurements were used for comparison of the field size dependence. Elsewhere, values from the linac monitoring chambers were used. Results In the range from 10 to 1000 Monitor Units (MU), the detector was linear within 1 % for all combinations of gain settings and energies. The dose rate dependence was also within 1% for all energies and for 2 out of 3 gain settings. Regarding field size dependency, the ratio of ionization chamber and panel values was 0.94 and 0.98 for the conventional EPID and GEMini respectively, at 20 x 20 cm2 and 10MV. For 6MV, 6MVFFF and 10MVFFF these ratios were 0.97, 0.98, 0.99 for the GEMini, and 0.95, 0.97 and 0.97 for the conventional EPID. Similar performance between the GEMini and conventional EPID is observed for field sizes smaller than 10 x 10 cm2. The transmission response was within 5 % for all energies for thicknesses up to 30 cm, compared to 10 – 20 % for a conventional EPID. The off‐axis response for shifts up to 16 cm was within 1 % and 3 % for 6 MV and 10 MV, with and without phantom. The rise and fall of the signal from the detector correspond well to monitor chamber measurements indicating little ghosting and image lag, regardless of gain setting. Conclusion The GEM EPID exhibits dose rate dependence and linearity within 1%, and negligible ghosting and image lag. In this regard, it performs particularly well using 50 V and 250 V of gain, and either could be chosen. For higher sensitivity 250 V is the recommended base gain setting, though other applications may warrant different gains. For most tests performed in this study, the GEM EPID demonstrates a more water‐equivalent response than conventional EPIDs making GEMs a viable technology for dosimetry in radiation therapy.Keywords
This publication has 19 references indexed in Scilit:
- Transit and non‐transit 3D EPID dosimetry versus detector arrays for patient specific QAJournal of Applied Clinical Medical Physics, 2019
- Characterization of the a-Si EPID in the unity MR-linac for dosimetric applicationsPhysics in Medicine & Biology, 2018
- Comparison of forward- and back-projection in vivo EPID dosimetry for VMAT treatment of the prostatePhysics in Medicine & Biology, 2018
- Pixel response‐based EPID dosimetry for patient specific QAJournal of Applied Clinical Medical Physics, 2016
- EPID based in vivo dosimetry system: clinical experience and resultsJournal of Applied Clinical Medical Physics, 2016
- Clinical practice and evaluation of electronic portal imaging device for VMAT quality assuranceMedical Dosimetry, 2013
- Basic Sciences of Nuclear Medicine2011
- Comparison of ghosting effects for three commercial EPIDsMedical Physics, 2006
- Accurate two‐dimensional IMRT verification using a back‐projection EPID dosimetry methodMedical Physics, 2006
- Radiation damage of amorphous silicon, thin‐film, field‐effect transistorsMedical Physics, 1996