Impact of Varying Amplitudes of Upward and Downward Motion on the Torque/Force Generation, Canal-Centering Ability, and Cyclic Fatigue Resistance of Nickel–Titanium Reciprocating Instrument
Open Access
- 31 December 2024
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 15 (1) , 288
Abstract
The optimal amplitude of pecking motion during nickel–titanium rotary root canal instrumentation remains unclear. This study evaluated how different pecking amplitudes influence the force/torque generation, canal-centering ability, and cyclic fatigue resistance of the Reciproc Blue R25 instrument. Thirty J-shaped resin canals were instrumented with Reciproc Blue using automated instrumentation and a torque/force sensing device with an up-and-down speed of 50 mm/min and varying pecking amplitudes: Group S (1 s apically, 0.5 s coronally), Group M (2 s apically, 1 s coronally), and Group L (4 s apically, 2 s coronally) (n = 10 each). Force/torque were recorded, and canal-centering ratios were determined. In a dynamic cyclic fatigue test, the Reciproc Blue instruments were rotated in a curved metal canal with 1 mm or 3 mm back-and-forth motions (n = 10 each). Group L exhibited significantly greater maximum downward force and cumulative upward (screw-in) force than Group S (p < 0.05), but torque was not significantly different across groups. Groups M and L exhibited a significantly smaller canal-centering ratio than Group S (p < 0.05). Pecking at 3 mm led to a significantly longer time to fracture and shorter fragment lengths than 1 mm pecking (p < 0.05). In the Reciproc Blue instrumentation, higher pecking amplitude improved canal-centering ability and cyclic fatigue resistance while increasing upward and downward forces.Keywords
Funding Information
- Japan Society for the Promotion of Science (21K16963ZA, 22H01041ZF, 22K17051ZA)
This publication has 47 references indexed in Scilit:
- The Role of Pecking Motion Depths in Dynamic Cyclic Fatigue Resistance: In Vitro StudyInternational Dental Journal, 2024
- Impact of pecking amplitude on cyclic fatigue of new nickel–titanium filesClinical and Experimental Dental Research, 2023
- Blue Thermomechanical Treatment Optimizes Fatigue Resistance and Flexibility of the Reciproc FilesJournal of Endodontics, 2017
- Kinematic Effects of Nickel-Titanium Instruments with Reciprocating or Continuous Rotation Motion: A Systematic Review of In Vitro StudiesJournal of Endodontics, 2016
- Effects of Pitch Length and Heat Treatment on the Mechanical Properties of the Glide Path Preparation InstrumentsJournal of Endodontics, 2016
- Shaping Ability of Different Nickel-Titanium Systems in Simulated S-shaped Canals with and without Glide PathJournal of Endodontics, 2014
- Differences in Cyclic Fatigue Resistance at Apical and Coronal Levels of Reciproc and WaveOne New FilesJournal of Endodontics, 2012
- A Retrospective Study of Endodontic Treatment Outcome between Nickel-Titanium Rotary and Stainless Steel Hand Filing TechniquesJournal of Endodontics, 2009
- A Comparison of the Canal Centering Ability of ProFile, K3, and RaCe Nickel Titanium Rotary SystemsJournal of Endodontics, 2006
- Rotary NiTi Instrument Fracture and its ConsequencesJournal of Endodontics, 2006