Comparison of anterior segment measurements with optical low-coherence reflectometry and rotating dual Scheimpflug analysis
Authors:
- 1 July 2014
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 40 (7) , 1170-1176
Abstract
To compare the anterior segment measurements obtained with optical low-coherence reflectometry (OLCR) (Lenstar LS 900, version 2.1.1) and rotating dual Scheimpflug analysis (Galilei G1, version 3). Private practice, Policlinic Lleida, Lleida, Spain. Prospective comparative observational study. The following measurements were performed in patients requiring a preoperative study for a refractive procedure or cataract surgery and healthy volunteers from the clinic's staff: central corneal thickness (CCT), anterior chamber depth (ACD), horizontal limbal distance (white-to-white distance [WTW]), pupil diameter, keratometry (K) readings at the steepest meridian (steep K) and the flattest meridian (flat K), corneal astigmatism power, and plus astigmatic cylinder. Measurements were taken using the OLCR device and rotating dual Scheimpflug analyzer. The main outcome measure was the degree of agreement between steep K, flat K, astigmatism power, cylinder axis, CCT, ACD, WTW, and pupil diameter measurements. The study enrolled 100 eyes of 100 subjects. Steep K, flat K, corneal astigmatism power, cylinder axis, and WTW measurements with the OLCR device and Scheimpflug analyzer showed narrow 95% limits of agreement (LoA), which implies good agreement (P>.05, Bland-Altman plot analysis). In contrast, the range and 95% LoA for CCT, ACD, and pupil diameter values were statistically significantly different (P<.05, Bland-Altman plot analysis). In clinical practice, the OLCR device and the rotating dual Scheimpflug analyzer system can be used interchangeably for WTW measurements and K readings but not for CCT, ACD, and pupil diameter values. No author has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 21 references indexed in Scilit:
- Evaluation of the repeatability of the Lenstar and comparison with two other non‐contact biometric devices in myopesClinical and Experimental Optometry, 2013
- Comparison of Central Corneal Thickness and Anterior Chamber Depth Measured Using LenStar LS900, Pentacam, and Visante AS-OCTCornea, 2012
- Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometryJournal of Cataract & Refractive Surgery, 2011
- Comparability and Intra-/Interobserver Reliability of Anterior Chamber Depth Measurements With the Pentacam and IOLMasterJournal of Refractive Surgery, 2008
- Comparison of Anterior Chamber Depth Measurements using Orbscan II and IOLMasterEuropean Journal of Ophthalmology, 2007
- Corneal Radii and Anterior Chamber Depth Measurements Using the IOLMaster Versus the PentacamJournal of Refractive Surgery, 2007
- Internal anterior chamber diameter using optical coherence tomography compared with white-to-white distances using automated measurementsJournal of Cataract & Refractive Surgery, 2006
- Anterior chamber depth measurements in phakic and pseudophakic eyes: Pentacam versus ultrasound deviceJournal of Cataract & Refractive Surgery, 2006
- Comparison of manual and automated methods to determine horizontal corneal diameterJournal of Cataract & Refractive Surgery, 2004
- Anterior chamber depth in relation to refractive status measured with the Orbscan II Topography SystemJournal of Cataract & Refractive Surgery, 2003