Air Damping Analysis of a Micro-Coriolis Mass Flow Sensor
Open Access
Abstract
A micro-Coriolis mass flow sensor is a resonating device that measures small mass flows of fluid. A large vibration amplitude is desired as the Coriolis forces due to mass flow and, accordingly, the signal-to-noise ratio, are directly proportional to the vibration amplitude. Therefore, it is important to maximize the quality factor Q so that a large vibration amplitude can be achieved without requiring high actuation voltages and high power consumption. This paper presents an investigation of the Q factor of different devices in different resonant modes. Q factors were measured both at atmospheric pressure and in vacuum. The measurement results are compared with theoretical predictions. In the atmospheric environment, the Q factor increases when the resonance frequency increases. When reducing the pressure from 1 bar to bar, the Q factor almost doubles. At even lower pressures, the Q factor is inversely proportional to the pressure until intrinsic effects start to dominate, resulting in a maximum Q factor of approximately 7200.
Keywords
Funding Information
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek (15026)
This publication has 29 references indexed in Scilit:
- Study of the out-of-plane vibrational modes in thin-film amorphous silicon micromechanical disk resonatorsJournal of Applied Physics, 2013
- Application of piezoelectric tuning forks in liquid viscosity and density measurementsSensors and Actuators B: Chemical, 2011
- In-plane acoustic reflectors for reducing effective anchor loss in lateral–extensional MEMS resonatorsJournal of Micromechanics and Microengineering, 2011
- Modeling, design, fabrication and characterization of a micro Coriolis mass flow sensorJournal of Micromechanics and Microengineering, 2010
- Investigation of the effects of air on the dynamic behavior of a small cantilever beamJournal of Sound and Vibration, 2007
- Squeeze film air damping in MEMSSensors and Actuators A: Physical, 2007
- Comments on the size effect on the microcantilever quality factor in free air spaceJournal of Micromechanics and Microengineering, 2006
- Silicon resonant microcantilevers for absolute pressure measurementJournal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena, 2006
- Theoretical and Experimental Study on Airflow Damping of Vibrating MicrocantileversJournal of Vibration and Acoustics, 1999
- Dependence of the quality factor of micromachined silicon beam resonators on pressure and geometryJournal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena, 1992