Motion‐ and Field‐Robust Mesoscopic Whole‐Brain T2*‐Weighted Imaging at 7 and 11.7 T Using Servo Navigation
Authors:
- Matthias SergerMSMatthias Serger
- Rüdiger StirnbergRSRüdiger Stirnberg
- Philipp EhsesPEPhilipp Ehses
- Malte RiedelMRMalte Riedel
- Thomas UlrichTUThomas Ulrich
- Caroline Le SterCSCaroline Le Ster
- Franck MauconduitFMFranck Mauconduit
- Vincent GrasVGVincent Gras
- Alexis AmadonAAAlexis Amadon
- Alexandre VignaudAVAlexandre Vignaud
- Son ChuSCSon Chu
- Shajan GunamonySGShajan Gunamony
Open Access
- 12 January 2026
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 95 (5) , 2658-2670
Abstract
Purpose: To mitigate artifacts related to motion and field changes in high‐resolution ‐weighted human brain imaging using servo navigation at ultra‐high fields up to 11.7 T. Methods: MR‐based servo navigators were integrated into a segmented 3D‐EPI sequence to allow for prospective correction of involuntary head motion and first‐order shim changes. Seven subjects were scanned with whole‐brain protocols at 0.3 mm isotropic resolution with and without correction at 7 and 11.7 T. Validation was performed on detailed brain vasculature in scans with involuntary motion. Results: Blurring of small veins was reduced by servo navigation for all subjects and across field strengths. In case of involuntary large motion, the method preserved image quality, while uncorrected motion led to severe artifacts. In case of microscopic motion, reduced blurring and shading in the frontal lobe demonstrate the additional benefit of prospective field drift correction. Conclusion: Servo‐navigated segmented 3D‐EPI improves 0.3 mm isotropic whole‐brain ‐weighted imaging under realistic motion and field changes within 5.5 to 11 min scan time at 11.7 and 7 T.Keywords
Funding Information
- Horizon 2020 Framework Programme (885876)
This publication has 59 references indexed in Scilit:
- Head motion detection using FID navigatorsMagnetic Resonance in Medicine, 2011
- Adaptive non‐local means denoising of MR images with spatially varying noise levelsJournal of Magnetic Resonance Imaging, 2009
- Navigator accuracy requirements for prospective motion correctionMagnetic Resonance in Medicine, 2009
- 3D cine displacement‐encoded MRI of pulsatile brain motionMagnetic Resonance in Medicine, 2008
- Real‐time shimming to compensate for respiration‐induced B0 fluctuationsMagnetic Resonance in Medicine, 2007
- Real‐time rigid body motion correction and shimming using cloverleaf navigatorsMagnetic Resonance in Medicine, 2006
- Spherical navigator echoes for full 3D rigid body motion measurement in MRIMagnetic Resonance in Medicine, 2001
- Adaptive reconstruction of phased array MR imageryMagnetic Resonance in Medicine, 2000
- Prospective multiaxial motion correction for fMRIMagnetic Resonance in Medicine, 2000
- Orbital navigator echoes for motion measurements in magnetic resonance imagingMagnetic Resonance in Medicine, 1995