On the Stability of Poly‐Ethylenedioxythiopene as Coating Material for Active Neural Implants
- 15 March 2011
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 35 (3) , 245-248
Abstract
This article deals with the stability of poly‐ethylenedioxythiopene (PEDOT) coatings under high loads of current pulses. Test parameters were chosen to match many peripheral nervous system applications in regard of charge injection, pulse width, and repetition frequency. PEDOT coatings were characterized with electrochemical impedance spectroscopes and pulse tests. After 60–100 million pulses, impedance increased and a more capacitive behavior was observed. A mean‐time‐to‐failure of 127 million pulses could be calculated, suggesting a stable coating for at least this time frame with superior properties in regard of conventional platinum electrodes.Keywords
This publication has 6 references indexed in Scilit:
- Double nerve intraneural interface implant on a human amputee for robotic hand controlClinical Neurophysiology, 2010
- Charge Storage: Stability measures in implantable electrodes2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2009
- Cell attachment functionality of bioactive conducting polymers for neural interfacesBiomaterials, 2009
- Neural Stimulation and Recording ElectrodesAnnual Review of Biomedical Engineering, 2008
- A voltage-controlled current source with regulated electrode bias-voltage for safe neural stimulationJournal of Neuroscience Methods, 2008
- Implantable biomedical microsystems for neural prosthesesIEEE Engineering in Medicine and Biology Magazine, 2005