Fast‐Response Variable‐Stiffness Magnetic Catheters for Minimally Invasive Surgery
Top Cited Papers
Open Access
- 15 January 2024
- journal article
- research article
- Published by Wiley in Advanced Science
- Vol. 11 (12) , e2305537
Abstract
In minimally invasive surgery, such as cardiac ablation, magnetically steered catheters made of variable‐stiffness materials can enable higher dexterity and higher force application to human tissue. However, the long transition time between soft and rigid states leads to a significant increase in procedure duration. Here, a fast‐response, multisegmented catheter is described for minimally invasive surgery made of variable‐stiffness thread (FRVST) that encapsulates a helical cooling channel. The rapid stiffness change in the FRVST, composed of a nontoxic shape memory polymer, is achieved by an active cooling system that pumps water through the helical channel. The FRVST displays a 66 times stiffness change and a 26 times transition enhancement compare with the noncooled version. The catheter allows for selective bending of each segment up to 127° in air and up to 76° in water under an 80 mT external magnetic field. The inner working channel can be used for cooling an ablation tip during a procedure and for information exchange via the deployment of wires or surgical tools.Keywords
Funding Information
- nccr – on the move (200020B 185039)
- Swiss Bridge Foundation (20B2‐1 18086)
This publication has 55 references indexed in Scilit:
- Use of Asymmetric Bidirectional Catheters with Different Curvature Radius for Catheter Ablation of Cardiac ArrhythmiasPacing and Clinical Electrophysiology, 2013
- Steerable Versus Nonsteerable Sheath Technology in Atrial Fibrillation AblationCirculation: Arrhythmia and Electrophysiology, 2011
- Percutaneous Cell Delivery into the Heart Using Hydrogels Polymerizing in SituCell Transplantation, 2009
- Dip coating assisted polylactic acid deposition on steel surface: Film thickness affected by drag force and gravityMaterials Letters, 2008
- Experience of robotic catheter ablation in humans using a novel remotely steerable catheter sheathJournal of Interventional Cardiac Electrophysiology, 2008
- Polyurethane Shape‐Memory Polymers Demonstrate Functional Biocompatibility In VitroMacromolecular Bioscience, 2007
- Remote-controlled catheter ablation of accessory pathways: results from the magnetic laboratoryEuropean Heart Journal, 2006
- In vitro interaction of human fibroblasts and platelets with a shape‐memory polyurethaneJournal of Biomedical Materials Research Part A, 2005
- In Vivo Measurement of Swine Endocardial Convective Heat Transfer CoefficientIEEE Transactions on Biomedical Engineering, 2004
- Influence of applied stress on mitotic response of arteries to injury with a balloon catheter: quantitative study in rat thoracic aortaCardiovascular Research, 1989