Design and Validation of a High-Sensitivity FBG-Based Triaxial Force Sensor With Serially Configured Flexures for Percutaneous Puncture
- 7 November 2025
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 74 (00189456) , 1-13
Abstract
This article introduces an innovative fiber Bragg grating (FBG)-based triaxial force sensor with high sensitivity for percutaneous puncture applications. The proposed sensor employs the serially configured axial and radial force-sensitive flexures, incorporating tightly suspended optical fibers inscribed with FBG elements. Leveraging the rigid-body replacement method, the designed flexures exhibit selective sensitivities to axial and radial loads, enabling low-coupling triaxial force sensing. The axial force-sensitive flexure combined a semi-bridge mechanism with symmetrical L-shaped levers, amplifying output force while reversing deformation direction. This design maintains the measuring fiber in a purely tensile state, obviating the need for preload adhesion, thereby streamlining assembly and augmenting fiber durability. Finite element method (FEM)-based simulation and structural parameter optimization have been implemented to further enhance the sensitivity. The sensor prototype has been calibrated employing a nonlinear decoupling algorithm based on a backpropagation neural network (BPNN), achieving high resolutions of 2.6 mN within the axial range of 8 N and 0.25 mN within the radial range of ±1 N, with the crosstalk limited to 4.36%. Puncture experiments demonstrated a root mean square error (RMSE) of 3-D resultant forces not exceeding 12.3 mN on ex-vivo tissues, validating the sensor's efficacy and potential in robotic puncture surgery.Keywords
Funding Information
- Research Project of Tianjin Science and Technology-Tianjin Natural Science Foundation (24JCYBJC01350)
- National Natural Science Foundation of China (92148201, 52475029)
- International Institute for Innovative Design and Intelligent Manufacturing of Tianjin University in Zhejiang, Shaoxing, China
This publication has 50 references indexed in Scilit:
- Robot-Assisted CT-Guided Biopsy with an Artificial Intelligence–Based Needle-Path Generator: An Experimental Evaluation Using a Phantom ModelJournal of Vascular and Interventional Radiology, 2025
- A Review of Recent Advancements in Sensor‐Integrated Medical ToolsAdvanced Science, 2024
- Flexible Biopsy Robot with Force Sensing for Deep Lung ExaminationAdvanced Intelligent Systems, 2023
- An Update of In Vivo Application of Artificial Intelligence and Robotics for Percutaneous Nephrolithotripsy: Results from a Systematic ReviewCurrent Urology Reports, 2023
- Percutaneous Kidney Puncture with Three-dimensional Mixed-reality Hologram Guidance: From Preoperative Planning to Intraoperative NavigationEuropean Urology, 2022
- A robotic puncture system with optical and mechanical feedback under respiratory motionThe International Journal of Medical Robotics and Computer Assisted Surgery, 2022
- A novel multi-DoF surgical robotic system for brachytherapy on liver tumor: Design and controlInternational Journal of Computer Assisted Radiology and Surgery, 2021
- Ultrasound-guided needle insertion robotic system for percutaneous punctureInternational Journal of Computer Assisted Radiology and Surgery, 2021
- A novel approach to an automated needle insertion in brachytherapy proceduresMedical & Biological Engineering & Computing, 2017
- A combination of experimental and finite element analyses of needle–tissue interaction to compute the stresses and deformations during injection at different anglesJournal of Clinical Monitoring and Computing, 2015