The “collimator monitoring fill factor” of a two‐dimensional detector array, a measure of its ability to detect collimation errors
- 17 January 2017
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 44 (3) , 1128-1138
Abstract
Purpose: Two‐dimensional detector arrays are routinely used for constancy checks and treatment plan verification in photon‐beam radiotherapy. In addition to the spatial resolution of the dose profiles, the “coverage” of the radiation field with respect to the detection of any beam collimation deficiency appears as the second characteristic feature of a detector array. The here proposed “collimator monitoring fill factor” (CM fill factor) has been conceived to serve as a quantitative characteristic of this “coverage”.Methods: The CM fill factor is defined as the probability of a 2D array to detect any collimator position error. Therefore, it is represented by the ratio of the “sensitive area” of a single detector, in which collimator position errors are detectable, and the geometrical “cell area” associated with this detector within the array. Numerical values of the CM fill factor have been Monte Carlo simulated for 2D detector arrays equipped with air‐vented ionization chambers, liquid‐filled ionization chambers and diode detectors and were compared with the “FWHM fill factor” defined by Gago‐Arias et al. (2012).Results: For arrays with vented ionization chambers, the differences between the CM fill factor and the FWHM fill factor are moderate, but occasionally the latter exceeds unity. For narrower detectors such as liquid‐filled ionization chambers and Si diodes and for small sampling distances, large differences between the FWHM fill factor and the CM fill factor have been observed. These differences can be explained by the shapes of the fluence response functions of these narrow detectors.Conclusions: A new parameter “collimator monitoring fill factor” (CM fill factor), applicable to quantitate the collimator position error detection probability of a 2D detector array, has been proposed. It is designed as a help in classifying the clinical performance of two‐dimensional detector arrays in photon‐beam radiotherapy.Keywords
This publication has 19 references indexed in Scilit:
- Experimental determination of the lateral dose response functions of detectors to be applied in the measurement of narrow photon-beam dose profilesPhysics in Medicine & Biology, 2015
- Understanding the lateral dose response functions of high-resolution photon detectors by reverse Monte Carlo and deconvolution analysisPhysics in Medicine & Biology, 2015
- Measurement of spatial response functions of dosimetric detectorsPhysics in Medicine & Biology, 2015
- Implementing RapidArc into clinical routine: A comprehensive program from machine QA to TPS validation and patient QAMedical Physics, 2011
- Assessment of the setup dependence of detector response functions for mega‐voltage linear acceleratorsMedical Physics, 2010
- The extraction of true profiles for TPS commissioning and its impact on IMRT patient-specific QAMedical Physics, 2008
- Two‐dimensional ionization chamber arrays for IMRT plan verificationMedical Physics, 2006
- Characterization of a 2D ion chamber array for the verification of radiotherapy treatmentsPhysics in Medicine & Biology, 2005
- Evaluation of a 2D diode array for IMRT quality assuranceRadiotherapy and Oncology, 2004
- The volume effect of detectors in the dosimetry of small fields used in IMRTMedical Physics, 2003