Stability, survival, and tolerability of a 4.5-mm-wide bone-anchored hearing implant: 6-month data from a randomized controlled clinical trial
Authors:
Open Access
- 20 March 2015
- journal article
- research article
- Published by Springer Nature in European Archives of Oto-Rhino-Laryngology
- Vol. 273 (1) , 105-111
Abstract
The objective of this study was to compare the stability, survival, and tolerability of 2 percutaneous osseointegrated titanium implants for bone conduction hearing: a 4.5-mm diameter implant (test) and a 3.75-mm diameter implant (control). Fifty-seven adult patients were included in this randomized controlled clinical trial. Sixty implants were allocated in a 2:1 (test–control) ratio. Follow-up visits were scheduled at 7, 14, 21, and 28 days; 6 and 12 weeks; and 6 months. At every visit, implant stability quotient (ISQ) values were recorded by means of resonance frequency analysis (RFA) and skin reactions were evaluated according to the Holgers classification. Implants were loaded with the bone conduction device at 3 weeks. Hearing-related quality of life was evaluated using the Abbreviated Profile of Hearing Aid Benefit (APHAB), the Glasgow Benefit Inventory (GBI), and the Glasgow Health Status Inventory (GHSI). ISQ values were statistically significantly higher for the test implant compared to the control implant. No implants were lost and soft tissue reactions were comparable for both implants. Positive results were reported in the hearing-related quality of life questionnaires. These 6-month results indicate that both implants and their corresponding hearing devices are safe options for hearing rehabilitation in patients with the appropriate indications. Loading at 3 weeks did not affect the stability of either implant.Keywords
This publication has 25 references indexed in Scilit:
- Assessment of More Than 1,000 Implanted Percutaneous Bone Conduction DevicesOtology & Neurotology, 2012
- Stability, Survival, and Tolerability of a Novel Baha Implant SystemOtology & Neurotology, 2011
- An Overview of Different Systems: The Bone-Anchored Hearing Aid2011
- Benefit and Quality of Life in Older Bone-Anchored Hearing Aid UsersOtology & Neurotology, 2010
- Bone‐anchored hearing aid subjective benefit for unilateral deafnessThe Laryngoscope, 2010
- Clinical Outcome of the Simplified Surgical Technique for BAHA ImplantationOtology & Neurotology, 2008
- Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implicationsPeriodontology 2000, 2008
- Effect of implant size and shape on implant success rates: A literature reviewThe Journal of Prosthetic Dentistry, 2005
- The Abbreviated Profile of Hearing Aid BenefitEar & Hearing, 1995
- Clinical, immunological and bacteriological evaluation of adverse reactions to skin‐penetrating titanium implants in the head and neck regionContact Dermatitis, 1992