Abstract
Preclinical evaluation of total knee arthroplasty (TKA) components is essential to understanding their mechanical behavior and developing strategies for improving joint stability. While preclinical testing of TKA components has been useful in quantifying their effectiveness, such testing can be criticized for lacking clinical relevance, as the important contributions of surrounding soft tissues are either neglected or greatly simplified. The purpose of our study was to develop and determine if subject-specific virtual ligaments reproduce a similar behavior as native ligaments surrounding TKA joints. Six TKA knees were mounted to a motion simulator. Each was subjected to tests of anterior–posterior (AP), internal–external (IE), and varus–valgus (VV) laxity. The forces transmitted through major ligaments were measured using a sequential resection technique. By tuning the measured ligament forces and elongations to a generic nonlinear elastic ligament model, virtual ligaments were designed and used to simulate the soft tissue envelope around isolated TKA components. The average root-mean-square error (RMSE) between the laxity results of TKA joints with native versus virtual ligaments was 3.5 ± 1.8 mm during AP translation, 7.5 ± 4.2 deg during IE rotations, and 2.0 ± 1.2 deg during VV rotations. Interclass correlation coefficients (ICCs) indicated a good level of reliability for AP and IE laxity (0.85 and 0.84). To conclude, the advancement of virtual ligament envelopes as a more realistic representation of soft tissue constraint around TKA joints is a valuable approach for obtaining clinically relevant kinematics when testing TKA components on joint motion simulators.Keywords
This publication has 55 references indexed in Scilit:
- Knee replacementThe Lancet, 2012
- Implant Design Influences Tibial Post Wear Damage in Posterior-stabilized KneesClinical Orthopaedics and Related Research, 2011
- Effect of joint laxity on polyethylene wear in total knee replacementJournal of Biomechanics, 2010
- The Epidemiology of Revision Total Knee Arthroplasty in the United StatesClinical Orthopaedics and Related Research, 2010
- Measurements of constraint of total knee replacementJournal of Biomechanics, 2005
- Pattern of anterior cruciate ligament force in normal walkingJournal of Biomechanics, 2004
- Validation of a three-dimensional model of the kneeJournal of Biomechanics, 1996
- Articular contact in a three-dimensional model of the kneeJournal of Biomechanics, 1991
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Response of a two-dimensional dynamic model of the human knee to the externally applied forces and momentsJournal of Biomedical Engineering, 1983