Direct in-chamber radon-220 (thoron) emanation measurements for rare-event physics experiments
Open Access
- 1 March 2026
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 21 (03) , P03027
Abstract
Measuring radon emanation from detector materials is a key method for controlling radon, a significant background in rare-event physics experiments. Methods for measuring radon emanation are well-established but have predominantly focused on the 222 Rn isotope, the dominant radon isotope for these backgrounds. However, measurements of 220 Rn (thoron), the second most abundant radon isotope, remain relatively unexplored. 220 Rn emanation measurements are challenging because the 220 Rn must be transferred from the emanation chamber to the active detector within its short 55 s half-life. In this study, a direct in-chamber approach for measuring 220 Rn emanation is presented in which the sample is placed directly within the active detector chamber, thereby minimising losses during transfer. The method was demonstrated with a DURRIDGE RAD8 electrostatic radon detector, which measured 220 Rn emanation from low-activity thoriated rods with an activity of 76 ± 20 mBq. Compared with a conventional flowthrough 220 Rn emanation setup, the in-chamber method increased sensitivity by a factor of 3. Using helium as the carrier gas provided a further sensitivity increase, giving an overall sensitivity gain of ∼ 5. These results indicate that in-chamber 220 Rn emanation measurements provide an effective tool for low-background experiments and have the potential to accelerate radon studies by exploiting the shorter half-life of 220 Rn.Keywords
All Related Versions
This publication has 17 references indexed in Scilit:
- Ultra-sensitive radon assay using an electrostatic chamber in a recirculating systemNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2025
- Development of high-sensitivity radon emanation measurement systems with surface treatment optimizationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2025
- Enhancing material screening at Boulby Underground Laboratory with XIA UltraLo-1800 alpha particle detectorsJournal of Instrumentation, 2025
- A high-sensitivity radon emanation detector system for future low-background experimentsJournal of Instrumentation, 2024
- Measurement of radon emanation and impurity adsorption from argon gas using ultralow radioactive zeoliteJournal of Instrumentation, 2024
- Radon emanation techniques and measurements for LZAIP Conference Proceedings, 2023
- Test of low radioactive molecular sieves for radon filtration in SF6 gas-based rare-event physics experimentsJournal of Instrumentation, 2021
- Radon backgrounds in the DEAP-1 liquid-argon-based Dark Matter detectorAstroparticle Physics, 2015
- High sensitivity radon emanation measurementsApplied Radiation and Isotopes, 2009
- A high-sensitivity large volume cryogenic detector for radon in gasJournal of Radioanalytical and Nuclear Chemistry, 2008