The effect of the screw pull-out rate on cortical screw purchase in unreamed and reamed synthetic long bones
Authors:
- 8 October 2009
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 224 (3) , 503-513
Abstract
Orthopaedic fracture fixation constructs are typically mounted on to human long bones using cortical screws. Biomechanical studies are increasingly employing commercially available synthetic bones. The aim of this investigation was to examine the effect of the screw pull-out rate and canal reaming on the cortical bone screw purchase strength in synthetic bone. Cylinders made of synthetic material were used to simulate unreamed (foam-filled) and reamed (hollow) human long bone with an outer diameter of 35 mm and a cortex wall thickness of 4 mm. The unreamed and reamed cylinders each had 56 sites along their lengths into which orthopaedic cortical bone screws (major diameter, 3.5 mm) were inserted to engage both cortices. The 16 test groups ( n = 7 screw sites per group) had screws extracted at rates of 1 mm/min, 5 mm/min, 10 mm/min, 20 mm/min, 30 mm/min, 40 mm/min, 50 mm/min, and 60 mm/min. The failure force and failure stress increased and were highly linearly correlated with pull-out rate for reamed ( R2 = 0.60 and 0.60), but not for unreamed ( R2 = 0.00 and 0.00) specimens. The failure displacement and failure energy were relatively unchanged with pull-out rate, yielding low coefficients for unreamed ( R2 = 0.25 and 0.00) and reamed ( R2 = 0.27 and 0.00) groups. Unreamed versus reamed specimens were statistically different for failure force ( p = 0.000) and stress ( p = 0.000), but not for failure displacement ( p = 0.297) and energy (0.054< p<1.000). This is the first study to perform an extensive investigation of the screw pull-out rate in unreamed and reamed synthetic long bone.Keywords
This publication has 44 references indexed in Scilit:
- The Effect of Screw Pullout Rate on Screw Purchase in Synthetic Cancellous BoneJournal of Biomechanical Engineering, 2008
- Cortical Screw Pullout Strength and Effective Shear Stress in Synthetic Third Generation Composite FemursJournal of Biomechanical Engineering, 2006
- The Biomechanics of Human Femurs in Axial and Torsional Loading: Comparison of Finite Element Analysis, Human Cadaveric Femurs, and Synthetic FemursJournal of Biomechanical Engineering, 2006
- An In Vitro Biomechanical Comparison of the Insertion Variables and Pullout Mechanical Properties of AO 6.5‐mm Standard Cancellous and 7.3‐mm Self‐Tapping, Cannulated Bone Screws in Foal Femoral BoneVeterinary Surgery, 2004
- Factors Affecting the Pullout Strength of Cancellous Bone ScrewsJournal of Biomechanical Engineering, 1996
- Mechanical validation of whole bone composite femur modelsJournal of Biomechanics, 1996
- Fixation pull-out response measurement of bone screws and porous-surfaced postsJournal of Biomechanics, 1994
- Role of loads and prosthesis material properties on the mechanics of the proximal femur after total hip arthroplastyJournal of Orthopaedic Research, 1992
- Holding Power of Orthopedic Screws in Equine Third Metacarpal and Metatarsal Bones: Part II. Adult Horse BoneVeterinary Surgery, 1985
- The Holding Power of Orthopedic Screws In VivoClinical Orthopaedics and Related Research, 1975