MRI geometric distortion: Impact on tangential whole‐breast IMRT
Open Access
- 8 September 2016
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (5) , 7-19
Abstract
The purpose of this study was to determine the impact of magnetic resonance imaging (MRI) geometric distortions when using MRI for target delineation and planning for whole-breast, intensity-modulated radiotherapy (IMRT). Residual system distortions and combined systematic and patient-induced distortions are considered. This retrospective study investigated 18 patients who underwent whole-breast external beam radiotherapy, where both CT and MRIs were acquired for treatment planning. Distortion phantoms were imaged on two MRI systems, dedicated to radiotherapy planning (a wide, closed-bore 3T and an open-bore 1T). Patient scans were acquired on the 3T system. To simulate MRI-based planning, distortion maps representing residual system distortions were generated via deformable registration between phantom CT and MRIs. Patient CT images and structures were altered to match the residual system distortion measured by the phantoms on each scanner. The patient CTs were also registered to the corresponding patient MRI scans, to assess patient and residual system effects. Tangential IMRT plans were generated and optimized on each resulting CT dataset, then propagated to the original patient CT space. The resulting dose distributions were then evaluated with respect to the standard clinically acceptable DVH and visual assessment criteria. Maximum residual systematic distortion was measured to be 7.9 mm ( ) and 11.9 mm ( ) for the 3T and 1T scanners, respectively, which did not result in clinically unacceptable plans. Eight of the plans accounting for patient and systematic distortions were deemed clinically unacceptable when assessed on the original CT. For these plans, the mean difference in PTV (volume receiving 95% prescription dose) was and for right- and left-sided patients, respectively. Residual system distortions alone had minimal impact on the dosimetry for the two scanners investigated. The combination of MRI systematic and patient-related distortions can result in unacceptable dosimetry for whole-breast IMRT, a potential issue when considering MRI-only radiotherapy treatment planning. PACS number(s): 87.61.-c, 87.57.cp, 87.57.nj, 87.55.D-
Keywords
Funding Information
- Cancer Australia and The National Breast Cancer Foundation (1033237)
- Liverpool and Macarthur Cancer Therapy Centres (1 R01 CA204189-01A1 (CGH))
- NSW Cancer Institute Leaders Program (PM)
This publication has 29 references indexed in Scilit:
- Inter‐ and intra‐fraction motion during radiation therapy to the whole breast in the supine position: A systematic reviewJournal of Medical Imaging and Radiation Oncology, 2012
- Magnetic Resonance Imaging– Versus Computed Tomography–Based Target Volume Delineation of the Glandular Breast Tissue (Clinical Target Volume Breast) in Breast-Conserving Therapy: An Exploratory StudyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Effects of Setup Errors and Shape Changes on Breast RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Current technological clinical practice in breast radiotherapy; results of a survey in EORTC-Radiation Oncology Group affiliated institutionsRadiotherapy and Oncology, 2010
- Systematisation of spatial uncertainties for comparison between a MR and a CT-based radiotherapy workflow for prostate treatmentsRadiation Oncology, 2009
- Feasibility of using MRI alone for 3D Radiation Treatment Planning in Brain TumorsJapanese Journal of Clinical Oncology, 2007
- MRI-based treatment planning for radiotherapy: Dosimetric verification for prostate IMRTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Effects of geometric distortion in 0.2T MRI on radiotherapy treatment planning of prostate cancerRadiotherapy and Oncology, 2004
- Correction for geometric distortion in echo planar images from B0 field variationsMagnetic Resonance in Medicine, 1995
- A technique for accurate magnetic resonance imaging in the presence of field inhomogeneitiesIEEE Transactions on Medical Imaging, 1992