Measurement accuracy and Cerenkov removal for high performance, high spatial resolution scintillation dosimetry
Authors:
- Louis ArchambaultLALouis Archambault
- A. Sam BeddarA. Sam Beddar
- Luc GingrasLGLuc Gingras
- René RoyRRRené Roy
- Luc BeaulieuLBLuc Beaulieu
- 23 December 2005
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 33 (1) , 128-135
Abstract
With highly conformal radiation therapy techniques such as intensity‐modulated radiation therapy, radiosurgery, and tomotherapy becoming more common in clinical practice, the use of these narrow beams requires a higher level of precision in quality assurance and dosimetry. Plastic scintillators with their water equivalence, energy independence, and dose rate linearity have been shown to possess excellent qualities that suit the most complex and demanding radiation therapy treatment plans. The primary disadvantage of plastic scintillators is the presence of Cerenkov radiation generated in the light guide, which results in an undesired stem effect. Several techniques have been proposed to minimize this effect. In this study, we compared three such techniques—background subtraction, simple filtering, and chromatic removal—in terms of reproducibility and dose accuracy as gauges of their ability to remove the Cerenkov stem effect from the dose signal. The dosimeter used in this study comprised a plastic scintillating fiber probe, an optical fiber, and a color charge‐coupled device camera. The whole system was shown to be linear and the total light collected by the camera was reproducible to within 0.31% for integration time. Background subtraction and chromatic removal were both found to be suitable for precise dose evaluation, with average absolute dose discrepancies of 0.52% and 0.67%, respectively, from ion chamber values. Background subtraction required two optical fibers, but chromatic removal used only one, thereby preventing possible measurement artifacts when a strong dose gradient was perpendicular to the optical fiber. Our findings showed that a plastic scintillation dosimeter could be made free of the effect of Cerenkov radiation.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada (2621005)
This publication has 18 references indexed in Scilit:
- Monte Carlo calculations of the absorbed dose and energy dependence of plastic scintillatorsMedical Physics, 2005
- Plastic scintillation dosimetry for radiation therapy: minimizing capture of Cerenkov radiation noisePhysics in Medicine & Biology, 2004
- Plastic scintillation dosimetry: optimization of light collection efficiencyPhysics in Medicine & Biology, 2003
- The volume effect of detectors in the dosimetry of small fields used in IMRTMedical Physics, 2003
- A miniature "scintillator-fiberoptic-PMT" detector system for the dosimetry of small fields in stereotactic radiosurgeryIEEE Transactions on Nuclear Science, 2001
- Miniature scintillating detector for small field radiation therapyMedical Physics, 1999
- Quality assurance of serial tomotherapy for head and neck patient treatmentsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1998
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurementsPhysics in Medicine & Biology, 1992
- Cerenkov light generated in optical fibres and other light pipes irradiated by electron beamsPhysics in Medicine & Biology, 1992