Toward Improving Patient Safety and Surgeon Comfort in a Synergic Robot-Assisted Eye Surgery: A Comparative Study
Open Access
- 27 January 2020
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in 2013 IEEE/RSJ International Conference on Intelligent Robots and Systems
- Vol. 2019, 7075-7082
Abstract
When robotic assistance is present into vitreoretinal surgery, the surgeon will experience reduced sensory input that is otherwise derived from the tool's interaction with the eye wall (sclera). We speculate that disconnecting the surgeon from this sensory input may increase the risk of injury to the eye and affect the surgeon's usual technique. On the other hand, robot autonomous motion to enhance patient safety might inhibit the surgeons tool manipulation and diminish surgeon comfort with the procedure. In this study, to investigate the parameters of patient safety and surgeon comfort in a robot-assisted eye surgery, we implemented three different approaches designed to keep the scleral force in a safe range during a synergic eye manipulation task. To assess the surgeon comfort during these procedures, the amount of interference with the surgeons usual maneuvers has been analyzed by defining quantitative comfort metrics. The first two utilized scleral force control approaches are based on an adaptive force control method in which the robot actively counteracts any excessive force on the sclera. The third control method is based on a virtual fixture approach in which a virtual wall is created for the surgeon in the unsafe directions of manipulation. The performance of the utilized approaches was evaluated in user studies with two experienced retinal surgeons and the outcomes of the procedure were assessed using the defined safety and comfort metrics. Results of these analyses indicate the significance of the opted control paradigm on the outcome of a safe and comfortable robot-assisted eye surgery.Keywords
This publication has 23 references indexed in Scilit:
- The meaning of surgeon's comfort in robotic surgeryThe American Journal of Surgery, 2014
- A multi-function force sensing instrument for variable admittance robot control in retinal microsurgery2014
- The introduction of a new robot for assistance in ophthalmic surgery2013
- Auditory Force Feedback Substitution Improves Surgical Precision during Simulated Ophthalmic SurgeryInvestigative Ophthalmology & Visual Science, 2013
- Impact of Robotic Assistance on Precision of Vitreoretinal Surgical ProceduresPLOS ONE, 2013
- Micron: An Actively Stabilized Handheld Tool for MicrosurgeryIEEE Transactions on Robotics, 2011
- Robot-Assisted Vitreoretinal SurgeryOphthalmology, 2009
- Nasa-Task Load Index (NASA-TLX); 20 Years LaterProceedings of the Human Factors and Ergonomics Society Annual Meeting, 2006
- Adaptive force control of position/velocity controlled robots: theory and experimentIEEE Transactions on Robotics and Automation, 2002
- Physiological tremor amplitude during retinal microsurgery2002