Clinical evaluation of BD Veritor ™ SARS-CoV-2 point-of-care test performance compared to PCR-based testing and versus the Sofia ® 2 SARS Antigen point-of-care test
Authors:
- Stephen YoungSYStephen Young
- Stephanie N. TaylorSTStephanie N. Taylor
- Catherine L. CammarataCCCatherine L. Cammarata
- Celine Roger-DalbertCRCeline Roger-Dalbert
- Amanda MontanoAMAmanda Montano
- Christen Griego-FullbrightCGChristen Griego-Fullbright
- Cameron BurgardCBCameron Burgard
- Catherine FernandezCFCatherine Fernandez
- Karen EckertKEKaren Eckert
- Jeffrey C. AndrewsJAJeffrey C. Andrews
- Huimiao RenHRHuimiao Ren
Preprint
Abstract
Objectives: The clinical performance of the BD Veritor™ System for Rapid Detection of SARS-CoV-2 antigen (Veritor), a chromatographic immunoassay that detects the SARS-CoV-2 nucleocapsid antigen as a point-of-care test, was evaluated on nasal specimens from individuals with COVID-19 symptoms. Methods and Materials: Two studies were completed to determine clinical performance. In the first study, nasal specimens and either nasopharyngeal or oropharyngeal specimens from 251 participants with COVID-19 symptoms (≤7 days from symptom onset [DSO]), ≥18 years of age, were utilized to compare Veritor with the Lyra ® SARS-CoV-2 PCR Assay (Lyra). In the second study, nasal specimens from 361 participants with COVID-19 symptoms (≤5 DSO), ≥18 years of age, were utilized to compare performance of Veritor to that of the Sofia ® 2 SARS Antigen FIA test (Sofia 2). Positive, negative, and overall percent agreement (PPA, NPA, and OPA, respectively) were the primary outcomes. Results: In study 1, PPA for Veritor, compared to Lyra, ranged from 81.8%-87.5% for 0-1 through 0-6 DSO ranges. In study 2, Veritor had a PPA, NPA, and OPA of 97.4%, 98.1%, and 98.1%, respectively, with Sofia 2. Discordant analysis showed one Lyra positive missed by Veritor and five Lyra positives missed by Sofia 2; one Veritor positive result was negative by Lyra. Conclusions: Veritor met FDA-EUA acceptance criteria for SARS-CoV-2 antigen testing (≥80% PPA point estimate) for the 0-5 and 0-6 DSO ranges. Veritor and Sofia 2 showed a high degree of agreement for SARS-CoV-2 detection. The Veritor test should facilitate rapid and reliable results for COVID-19 diagnosis utilizing easy-to-collect nasal swabs. Summary: The BD Veritor SARS-CoV-2 antigen test met FDA-EUA acceptance criteria for SARS-CoV-2 antigen testing for subjects with COVID-19 symptoms (0-5 days post-onset). BD Veritor and Quidel Sofia 2 antigen tests had good agreement for SARS-CoV-2 detection; discordant analysis favored Veritor.Keywords
All Related Versions
- Published version: Journal of Clinical Microbiology, 59 (1).
This publication has 21 references indexed in Scilit:
- Viral dynamics in mild and severe cases of COVID-19The Lancet Infectious Diseases, 2020
- Characteristics of and Important Lessons From the Coronavirus Disease 2019 (COVID-19) Outbreak in ChinaJAMA, 2020
- Diagnosis of Acute Respiratory Syndrome Coronavirus 2 Infection by Detection of Nucleocapsid ProteinmedRxiv, 2020
- Combination of RT‐qPCR testing and clinical features for diagnosis of COVID‐19 facilitates management of SARS‐CoV‐2 outbreakJournal of Medical Virology, 2020
- Laboratory readiness and response for novel coronavirus (2019-nCoV) in expert laboratories in 30 EU/EEA countries, January 2020Eurosurveillance, 2020
- Evaluating the accuracy of different respiratory specimens in the laboratory diagnosis and monitoring the viral shedding of 2019-nCoV infectionsmedRxiv, 2020
- STARD 2015: An Updated List of Essential Items for Reporting Diagnostic Accuracy StudiesRadiology, 2015
- Nucleocapsid Protein as Early Diagnostic Marker for SARSEmerging Infectious Diseases, 2004
- The use of “overall accuracy” to evaluate the validity of screening or diagnostic testsJournal of General Internal Medicine, 2004
- Characterization of a Novel Coronavirus Associated with Severe Acute Respiratory SyndromeScience, 2003