Technical assessment of a novel vertical CT system for upright radiotherapy simulation and treatment planning
Open Access
- 13 February 2026
- journal article
- Published by Wiley in Medical Physics
- Vol. 53 (2) , e70312
Abstract
Background: Upright patient positioning may provide anatomical advantages and more cost‐effective proton therapy using a fixed beamline. Purpose: To characterize image quality, imaging dose, and dose calculation accuracy for an upright CT scanner with a six‐degree‐of‐freedom patient positioning system. Methods: CT imaging dose (CTDI vol ) was measured at 120 kVp for head and thorax protocols. Image quality was evaluated using an ACR‐464 phantom. Mean CT number accuracy was assessed within inserts of known material, and uniformity as the difference in values at the center and periphery of uniform phantoms. High‐contrast resolution was assessed by visible line pairs and modulation transfer function (MTF). Low‐contrast performance was quantified by contrast‐to‐noise‐ratio (CNR). Spatial integrity was evaluated between fiducials 100 mm apart. Hounsfield unit to mass density and stopping‐power‐ratio calibrations were performed. Proton and photon plans were optimized on upright CT scans of a thorax phantom in heterogeneous and homogeneous regions. Dose was forward computed on a registered recumbent CT scan and agreement evaluated using 3D gamma analysis. Results: CTDI vol was 23.5 ± 0.02 mGy for the 16 cm head and 10.1 ± 0.01 mGy for the 32 cm body phantoms. Mean CT numbers (HU) were within the expected range for water (1.7) and acrylic (120.8). CT numbers were slightly [5–27 HU] out‐of‐range for air (−950.4), polyethylene (−78.8), and bone (823.0). Image uniformity was 20.2 HU and 35.0 HU for 20 and 48 cm diameter phantoms, respectively. Eight high‐contrast line pairs were visualized. The MTF equaled 4.4 cm −1 at 50% and 7.1 cm −1 at 10%. The median CNR was 0.93, below the ≥1.0 tolerance. Spatial integrity was ≤0.36 mm. Gamma pass rates were ≥99.8% for photon and ≥90.6% for proton plans with 1%/1 mm criteria, and ≥98.0% for all plans with 3%/2 mm criteria. Conclusion: Upright CT image quality and dose calculation accuracy are acceptable for photon and proton radiotherapy.Keywords
All Related Versions
This publication has 23 references indexed in Scilit:
- IPEM topical report: the first UK survey of dose indices from radiotherapy treatment planning computed tomography scans for adult patientsPhysics in Medicine & Biology, 2018
- Tolerance limits and methodologies forIMRTmeasurement‐based verificationQA:Recommendations ofAAPMTask Group No. 218Medical Physics, 2018
- Phase II Study of Proton-Based Stereotactic Body Radiation Therapy for Liver Metastases: Importance of Tumor GenotypeJNCI: Journal of the National Cancer Institute, 2017
- Consensus guidelines for postoperative stereotactic body radiation therapy for spinal metastases: results of an international surveyJournal of Neurosurgery: Spine, 2017
- Long-term outcomes after proton therapy, with concurrent chemotherapy, for stage II–III inoperable non-small cell lung cancerRadiotherapy and Oncology, 2015
- Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environmentJournal of Applied Clinical Medical Physics, 2014
- A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise‐power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantomMedical Physics, 2013
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66Medical Physics, 2003
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998