Liquid Blood Phantoms to Validate NIRS Oximeters: Yeast Versus Nitrogen for Deoxygenation
- 30 August 2018
- book chapter
- Published by Springer Nature
- Vol. 1072, 381-385
Abstract
Liquid blood phantoms are a tool to calibrate, test and compare near-infrared spectroscopy (NIRS) oximeters. They comprise a mixture of saline, blood and Intralipid, which is subsequently oxygenated and deoxygenated to assess the entire range of tissue oxygen saturation (StO2) from 0% to 100%. The aim was to investigate two different deoxygenation methods: yeast versus nitrogen (N2) bubbling. The phantom was oxygenated with pure O2 in both experiments, but deoxygenated by bubbling N2 in the first and by addition of yeast and glucose in the second experiment. A frequency domain NIRS instrument (OxiplexTS) was used as reference and to monitor changes in the reduced scattering coefficient (μs’) of the phantom. Both deoxygenation methods yielded comparable StO2 values. The deoxygenation was slower by a factor 2.8 and μs’ decreased faster when bubbling N2. The constant bubbling of N2 mechanically stresses the Intralipid emulsion and causes a decrease in μs’, probably due to aggregation of lipid droplets. Deoxygenation by N2 requires a more complex, air tight phantom. The gas flow cools the liquid and temperature needs to be monitored more closely. Consequently, we recommend using yeast for phantom deoxygenation.Keywords
This publication has 8 references indexed in Scilit:
- Quantifying the effect of adipose tissue in muscle oximetry by near infrared spectroscopyBiomedical Optics Express, 2016
- Comparison of tissue oximeters on a liquid phantom with adjustable optical propertiesBiomedical Optics Express, 2016
- Comparison of Near-Infrared Oximeters in a Liquid Optical Phantom with Varying Intralipid and Blood Content2016
- Cerebral oxygenation after birth – a comparison of INVOS ® and FORE ‐ SIGHT ™ near‐infrared spectroscopy oximetersActa Paediatrica, 2014
- A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodologyNeuroImage, 2014
- Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantomBiomedical Optics Express, 2013
- A Review of near Infrared Spectroscopy for Term and Preterm NewbornsJournal of Near Infrared Spectroscopy, 2012
- Monitoring Neonatal Regional Cerebral Oxygen Saturation in Clinical Practice: Value and PitfallsNeonatology, 2008