A novel Cherenkov radiation removal method for plastic scintillator detectors in a 0.35 T MR‐Linac
Open Access
- 15 August 2025
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 26 (8) , e70202
Abstract
Purpose: This study evaluates methods for removing Cherenkov radiation (CR) from plastic scintillation detectors (PSDs), focusing on constraints specific to a 0.35 T MR‐Linac system.Methods: Five CR‐removal methods were examined: cross calibration, fiber alone, multiloop, collimator rotation, and couch rotation. The first three (cross calibration, fiber alone, and multiloop) were tested on a 0.35 T MR‐Linac (ViewRay Inc., USA) using the BluePhysics PSD (Blue Physics LLC, USA). These methods do not require collimator or couch rotation. The remaining two methods (collimator rotation and couch rotation) were tested on a Varian TrueBeam (Varian Medical Systems, USA) for comparison. Measurements were performed under various setup configurations, and Cherenkov radiation extraction (CRE) values were calculated to determine each method's effectiveness.Results: The multiloop approach yielded a CRE of 0.7288, making it the most practical and robust for MR‐Linac constraints because it requires neither collimator nor couch rotation. The cross calibration and fiber alone methods produced CRE values of 0.7318 and 0.7569, respectively. Collimator rotation gave 0.7255, comparable to multiloop. In contrast, couch rotation resulted in 0.7489 but exhibited more variability, suggesting lower reliability.Conclusion: The multiloop method emerged as the most practical and robust technique for CR removal in 0.35 T MR‐Linac systems. Its simplicity and compatibility with MR‐Linac design constraints make it a highly effective approach for CR removal in PSD‐based radiotherapy applications.Keywords
This publication has 33 references indexed in Scilit:
- On the nature of the light produced within PMMA optical light guides in scintillation fiber-optic dosimetryPhysics in Medicine & Biology, 2013
- Three-dimensional Čerenkov tomography of energy deposition from ionizing radiation beamsOptics Letters, 2013
- Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: A comparison study of calibration procedures and validation in Cerenkov light‐dominated situationsMedical Physics, 2011
- Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamberPhysics in Medicine & Biology, 2009
- Accelerator beam data commissioning equipment and procedures: Report of the TG‐106 of the Therapy Physics Committee of the AAPMMedical Physics, 2008
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose deposition in a transverse magnetic fieldPhysics in Medicine & Biology, 2004
- A temporal method of avoiding the Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beamsPhysics in Medicine & Biology, 2002
- A new scintillator detector system for the quality assurance of60Co and high-energy therapy machinesPhysics in Medicine & Biology, 1994
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurementsPhysics in Medicine & Biology, 1992