The Effect of Poly Sterilization on Wear, Osteolysis and Survivorship of a Press-fit Cup at 10-Year Followup
Authors:
- Charles A. EnghCECharles A. Engh
- Robert H. HopperRHRobert H. Hopper
- Anderson C. EnghAEAnderson C. Engh
Open Access
- 1 February 2012
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 470 (2) , 462-470
Abstract
During the mid-1990s when our institution was using a press-fit porous-coated cup without supplemental initial fixation for primary THA, the manufacturer transitioned from gamma irradiation to gas plasma for the terminal sterilization of their polyethylene liners. At minimum 10-year followup, we asked whether the fixation achieved by solely relying on a press-fit would be durable and how different liner sterilization methods affected radiographic wear, osteolysis, and survivorship. We retrospectively reviewed 373 patients who underwent 398 primary THAs with a press-fit porous-coated cup between March 1995 and December 1996. Mean age at time of surgery was 61.5 ± 13.3 years and mean followup was 10.4 ± 3.7 years. We determined reasons for revision, survivorship, femoral head penetration, osteolysis, and wear-related complications. Among 20 revisions involving any component, seven were associated with wear and osteolysis. Kaplan-Meier survivorship, using component revision for any reason as an end point, was 95.7% (95% confidence interval, 93.6%–97.9%) at 10 years. Noncrosslinked liners sterilized with gas plasma demonstrated a mean head penetration rate of 0.20 ± 0.09 mm/year compared with 0.13 ± 0.07 mm/year for liners sterilized with gamma irradiation in air and 0.09 ± 0.04 mm/year for liners sterilized with gamma-irradiation with barrier packaging without oxygen. THAs with increased volumetric wear tended to demonstrate larger osteolytic lesions (r = 0.40) and there tended to be less osteolysis among the liners sterilized with gamma-irradiation with barrier packaging without oxygen. However, there was no difference in survivorship among the sterilization groups and there has been no cup or stem loosening associated with osteolysis. Durable biologic fixation through 10-year followup can be achieved by solely relying on an initial press-fit. Noncrosslinking gas plasma for terminal sterilization of the polyethylene liners was associated with greater head penetration rate than gamma irradiation. Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.Keywords
This publication has 44 references indexed in Scilit:
- Second-generation Extensively Porous-coated THA Stems at Minimum 10-year FollowupClinical Orthopaedics and Related Research, 2009
- Gamma Inert Sterilization: A Solution to Polyethylene Oxidation?Journal of Bone and Joint Surgery, 2009
- Can the Volume of Pelvic Osteolysis be Calculated without Using Computed Tomography?Clinical Orthopaedics and Related Research, 2009
- Patient Sensitivity to Polyethylene Particles with Cemented Total Hip ArthroplastyThe Journal of Arthroplasty, 2007
- Migration of the Duraloc cup after 5 yearsInternational Orthopaedics, 2007
- Second-generation Porous-coated Cementless Total Hip Arthroplasties Have High SurvivalClinical Orthopaedics and Related Research, 2006
- Oxidation and oxidation potential in contemporary packaging for polyethylene total joint replacement componentsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2006
- Increase in early polyethylene wear after sterilization with ethylene oxideRadiostereometric analyses of 201 total hipsActa Orthopaedica, 2003
- Osteogenesis imperfecta – where next?The Journal of Bone and Joint Surgery. British volume, 1997
- Nonparametric Estimation from Incomplete ObservationsJournal of the American Statistical Association, 1958