Electrocardiogram‐Synchronized Rotational Speed Change Mode in Rotary Pumps Could Improve Pulsatility
- 25 May 2011
- journal article
- Published by Wiley in Artificial Organs
- Vol. 35 (10) , 941-947
Abstract
Continuous‐flow left ventricular assist devices (LVADs) have greatly improved the prognosis of patients with end‐stage heart failure, even if continuous flow is different from physiological flow in that it has less pulsatility. A novel pump controller of continuous‐flow LVADs has been developed, which can change its rotational speed (RS) in synchronization with the native cardiac cycle, and we speculated that pulsatile mode, which increases RS just in the systolic phase, can create more pulsatility than the current system with constant RS does. The purpose of the present study is to evaluate the effect of this pulsatile mode of continuous‐flow LVADs on pulsatility in in vivo settings. Experiments were performed on eight adult goats (61.7 ± 7.5 kg). A centrifugal pump, EVAHEART (Sun Medical Technology Research Corporation, Nagano, Japan), was installed by the apex drainage and the descending aortic perfusion. A pacing lead for the detection of ventricular electrocardiogram was sutured on the anterior wall of the right ventricle. In the present study, we compared pulse pressure or other parameters in the following three conditions, including Circuit‐Clamp (i.e., no pump support), Continuous mode (constant RS), and Pulsatile mode (increase RS in systole). Assist rate was calculated by dividing pump flow (PF) by the sum of PF and ascending aortic flow (AoF). In continuous and pulsatile modes, these assist rates were adjusted around 80–90%. The following three parameters were used to evaluate pulsatility, including pulse pressure, dp/dt of aortic pressure (AoP), and energy equivalent pulse pressure (EEP = (∫PF*AoP dt)/(∫PF dt), mm Hg). The percent difference between EEP and mean AoP is used as an indicator of pulsatility, and normally it is around 10% of mean AoP in physiological pulse. Both pulse pressure and mean dp/dt max were decreased in continuous mode compared with clamp condition, while those were regained by pulsatile mode nearly to clamp condition (pulse pressure, clamp/continuous/pulsatile, 25.0 ± 7.6/11.7 ± 6.4/22.6 ± 9.8 mm Hg, mean dp/dt max, 481.9 ± 207.6/75.6 ± 36.2/351.1 ± 137.8 mm Hg/s, respectively). In clamp condition, %EEP was 10% higher than mean AoP (P = 0.0078), while in continuous mode, %EEP was nearly equivalent to mean AoP (N.S.). In pulsatile mode, %EEP was 9% higher than mean AoP (P = 0.038). Our newly developed pulsatile mode of continuous‐flow LVADs can produce pulsatility comparable to physiological pulsatile flow. Further investigation on the effect of this novel drive mode on organ perfusion is currently ongoing.Keywords
This publication has 30 references indexed in Scilit:
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceNew England Journal of Medicine, 2009
- Effects of Centrifugal, Axial, and Pulsatile Left Ventricular Assist Device Support on End-Organ Function in Heart Failure PatientsThe Journal of Heart and Lung Transplantation, 2009
- Differences in Pulsatile and Non-pulsatile Mechanical Circulatory Support in Long-Term UseThe Journal of Heart and Lung Transplantation, 2008
- Clinical outcome and bridge to transplant rate of left ventricular assist device recipient patients: comparison between continuous-flow and pulsatile-flow devices☆European Journal of Cardio-Thoracic Surgery, 2008
- Renal Function After Implantation of Continuous Versus Pulsatile Flow Left Ventricular Assist DevicesThe Journal of Heart and Lung Transplantation, 2008
- Effects of Pulsatile- and Continuous-flow Left Ventricular Assist Devices on Left Ventricular UnloadingThe Journal of Heart and Lung Transplantation, 2008
- End-organ Function in Patients on Long-term Circulatory Support With Continuous- or Pulsatile-flow Assist DevicesThe Journal of Heart and Lung Transplantation, 2007
- Clinical Outcomes Are Similar in Pulsatile and Nonpulsatile Left Ventricular Assist Device RecipientsThe Annals of Thoracic Surgery, 2007
- Myths and Truths of Pulsatile and Nonpulsatile Perfusion During Acute and Chronic Cardiac SupportArtificial Organs, 2004
- Left ventricular pressure and volume unloading during pulsatile versus nonpulsatile left ventricular assist device supportThe Annals of Thoracic Surgery, 2004