Abstract
Closed-loop charge balancers make a considerable contribution to a safe and secure electrical stimulation and additionally present an alternative to bulky blocking capacitors. This CMOS-integrated, 22 V compliant Twin-Track active charge balancer accomplishes both instantaneous and long-term balanced conditions. Long-term charge balancing is achieved by a cause-based PI-controlled offset compensation of the remaining electrode voltage. A Gm of only 1.5 nS offers a large (8 ms) time constant with only a 12 pF capacitor and leads to the first integrated PI-controller for offset compensation in neural implants. Instantaneous autonomous balancing is realized by a second concept, the consequence-based Inter-Pulse Charge Control. It uses a class-B architecture to supply the compensation currents. Thus, no additional references are required to define the safety window of the remaining electrode voltage. It consumes 31.8 μW in the idle state, while the delivered output power can be as high as 11 mW. The two complementary approaches can be combined to one system, providing the opportunity to achieve a better performance. The system consumes 56 μW and features a 36% mismatch correction in biphasic stimuli amplitude, as well as an autonomous 100% charge removal at maximum compensation amplitudes of 500 μA, which makes it a suitable complement to monophasic stimulators.Keywords
Funding Information
- German Research Foundation (DFG) (MA 2193/20-1)
- BrainLinks-BrainTools Cluster of Excellence (DFG) (EXC 1086)
This publication has 22 references indexed in Scilit:
- Blood pressure control with selective vagal nerve stimulation and minimal side effectsJournal of Neural Engineering, 2014
- BaroLoop: Using a multichannel cuff electrode and selective stimulation to reduce blood pressure2013
- Safety Ensuring Retinal Prosthesis With Precise Charge Balance and Low Power ConsumptionIEEE Transactions on Biomedical Circuits and Systems, 2013
- The Electrochemistry of Charge Injection at the Electrode/Tissue Interface2010
- Vagal afferent innervation and remodeling in the aortic arch of young-adult fischer 344 rats following chronic intermittent hypoxiaNeuroscience, 2009
- Charge Balancing in Functional Electrical Stimulators: A Comparative Study2007
- Switched-capacitor based implantable low-power wireless microstimulating systems2006
- Electrical stimulation of excitable tissue: design of efficacious and safe protocolsJournal of Neuroscience Methods, 2005
- Transconductance amplifier structures with very small transconductances: a comparative design approachIEEE Journal of Solid-State Circuits, 2002
- Chronic Electrical Stimulation of the Auditory Nerve using Non-charge-balanced StimuliActa Oto-Laryngologica, 1999