Optimising proton stopping power ratio prediction with spectral cone-beam CT
Open Access
- 16 July 2025
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (14) , 145023
Abstract
Objective. Cone-beam computed tomography (CBCT) is used for patient positioning in proton therapy, but not directly for treatment planning due to its inferior image quality compared to fan-beam CT. One way to improve its value for proton radiotherapy might be to use CBCT setups capable of extracting spectral information, which can be realised through several hardware configurations. Here, we compare different setups w.r.t. to their capability of predicting proton stopping power ratios (SPRs). Approach. We investigate six different spectral CBCT realisations in a simulation study, namely a single-source setup with either a dual-layer detector or a photon-counting detector (PCD), a kVp-switching setup with either an energy-integrating detector (EID) or a PCD, and a dual-source setup with either EIDs or PCDs. Our figure of merit is the normalised Cramér–Rao Lower Bound (nCRLB) on SPR variance based on projection data. We take (cross)scatter into account, and compare ideal and realistic detector models to help guide future detector developments. Each setup is optimised w.r.t. source spectra, mAs ratios and energy bin settings (where applicable). Main results. Assuming a realistic detector response, setups with a kVp-switching source perform best, with the setup paired with an EID slightly outperforming the PCD-based setup (nCRLBs of 2.74 and 2.81, respectively). However, if the mAs ratio of the kVp-switching source is fixed, the performance of the kVp-switching setup with an EID is significantly degraded (nCRLB = 9.46) and outperformed by PCD-based setups, with nCRLBs of 3.27, 3.45 and 3.60 for the dual-source setup with two PCDs, the single-source setup and the kVp-switching setup with one PCD, respectively. Spectra with higher source voltage or wider spectral separation generally yield lower CRLB values, and avoiding the spectral distortion caused by charge sharing in direct-conversion PCDs promises to lower CRLB values by about a third. Significance. We present an extensive comparison of spectral CBCT setups for their application in proton radiotherapy, using a methodology that allows to compare their theoretical limit of performance without being influenced by the choice of reconstruction algorithm or the conversion scheme from Hounsfield units to SPR values.Keywords
Funding Information
- Varian, a Siemens Healthineers company (2018016)
This publication has 55 references indexed in Scilit:
- Imaging performance of a LaBr3:Ce scintillation detector for photon counting x‐ray computed tomography: Simulation studyMedical Physics, 2024
- The technical development of photon-counting detector CTEuropean Radiology, 2023
- Silicon photomultiplier‐based scintillation detectors for photon‐counting CT: A feasibility studyMedical Physics, 2021
- Evaluation of carbon nanotube x‐ray source array for stationary head computed tomographyMedical Physics, 2020
- Markerless tumor tracking using fast‐kV switching dual‐energy fluoroscopy on a benchtop systemMedical Physics, 2019
- Current state and future applications of radiological image guidance for particle therapyMedical Physics, 2018
- Spatio‐energetic cross‐talk in photon counting detectors: Numerical detector model (PcTK) and workflow for CT image quality assessmentMedical Physics, 2018
- Modeling the frequency‐dependent detective quantum efficiency of photon‐counting x‐ray detectorsMedical Physics, 2017
- The potential of dual-energy CT to reduce proton beam range uncertaintiesMedical Physics, 2017
- Evaluation of the collision stopping power of elements and compounds for electrons and positronsThe International Journal of Applied Radiation and Isotopes, 1982