Digital magnification mammography with matched incident exposure: physical imaging properties and detectability of simulated microcalcifications
Authors:
- 18 March 2011
- journal article
- research article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 4 (2) , 156-163
Abstract
Our purpose was to evaluate the usefulness of digital magnification mammography with matched incident exposure by investigating the physical imaging properties and doing an observer performance test. A computed radiography system and a mammographic unit were used in this study. Contact and magnification radiographies of 1.2–1.8 in combination with focal spot sizes of 0.1 mm without grid and 0.3 mm with grid were performed. Physical imaging properties, namely, scatter fraction, total modulation transfer function (MTF) including the presampled MTF and the MTF of focal spot size, and Wiener spectrum (WS), were measured. Detail visibility was evaluated by use of free-response receiver operating characteristic analysis of the detectability of simulated microcalcifications. Scatter fractions decreased considerably as the magnification factor increased without grid technique. In the grid technique, scatter fractions for all magnification techniques were comparable. The total MTFs of magnification techniques with a focal spot size of 0.1 mm improved significantly compared with the conventional contact technique. However, the improvement of the total MTFs of magnification techniques with the combination of 0.3 mm focal spot size was small. The WSs degraded with an increase of the magnification factor compared with the contact technique due to the maintained exposure incident on the object. The observer performance test indicated that the 1.8 magnification technique with the 0.1 mm focal spot size provided higher detectability than did the contact technique. Digital magnification mammography under the same incident exposure conditions improved the detectability of microcalcifications.Keywords
This publication has 19 references indexed in Scilit:
- Noise Power SpectrumJapanese Journal of Radiological Technology, 2009
- Effect of Dose Reduction on the Ability of Digital Mammography to Detect Simulated MicrocalcificationsJournal of Digital Imaging, 2009
- Proposal and Verification of Presampled MTF Measurement by the Simple Analysis Method Using the Edge MethodJapanese Journal of Radiological Technology, 2008
- Imaging properties of digital magnification radiographyMedical Physics, 2006
- Breast dose from magnification films in mammographyBritish Journal of Radiology, 2005
- Maximum likelihood fitting of FROC curves under an initial‐detection‐and‐candidate‐analysis modelMedical Physics, 2002
- Comparison of low contrast detectability between a digital amorphous silicon and a screen‐film based imaging system for thoracic radiographyMedical Physics, 1999
- Physical and clinical evaluation of a 2,048 × 2,048-matrix image intensifier TV digital imaging system in bone radiographyAcademic Radiology, 1996
- Digital MammographyInvestigative Radiology, 1987
- The detection and significance of calcifications in the breast: a radiological and pathological studyBritish Journal of Radiology, 1976