Vectorcardiography as a Tool for Easy Optimization of Cardiac Resynchronization Therapy in Canine Left Bundle Branch Block Hearts
- 1 June 2012
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 5 (3) , 544-552
Abstract
Background—: In cardiac resynchronization therapy (CRT), optimization of left ventricular (LV) stimulation timing is often time consuming. We hypothesized that the QRS vector in the vectorcardiogram (VCG) reflects electric interventricular dyssynchrony, and that the QRS vector amplitude (VA QRS ), halfway between that during left bundle branch block (LBBB) and LV pacing, reflects optimal resynchronization, and can be used for easy optimization of CRT. Methods and Results—: In 24 canine hearts with LBBB (12 acute, 6 with heart failure, and 6 with myocardial infarction), the LV was paced over a wide range of atrioventricular (AV) delays. Surface ECGs were recorded from the limb leads, and VA QRS was calculated in the frontal plane. Mechanical interventricular dyssynchrony (MIVD) was determined as the time delay between upslopes of LV and right ventricular pressure curves, and systolic function was assessed as LV dP/dt max . VA QRS and MIVD were highly correlated ( r =0.94). The VA QRS halfway between that during LV pacing with short AV delay and intrinsic LBBB activation accurately predicted the optimal AV delay for LV pacing (1 ms; 95% CI, –5 to 8ms). Increase in LV dP/dt max at the VCG predicted AV delay was only slightly lower than the highest observed ∆LV dP/dt max (–2.7%; 95% CI, –3.6 to –1.8%). Inability to reach the halfway value of VA QRS during simultaneous biventricular pacing (53% of cases) was associated with suboptimal hemodynamic response, which could be corrected by sequential pacing. Conclusions—: The VA QRS reflects electric interventricular dyssynchrony and accurately predicts optimal timing of LV stimulation in canine LBBB hearts. Therefore, VCG may be useful as a reliable and easy tool for individual optimization of CRT.Keywords
This publication has 15 references indexed in Scilit:
- Primary Results From the SmartDelay Determined AV Optimization: A Comparison to Other AV Delay Methods Used in Cardiac Resynchronization Therapy (SMART-AV) TrialCirculation, 2010
- When is an optimization not an optimization? Evaluation of clinical implications of information content (signal-to-noise ratio) in optimization of cardiac resynchronization therapy, and how to measure and maximize itHeart Failure Reviews, 2010
- Myocardial Infarction Does Not Preclude Electrical and Hemodynamic Benefits of Cardiac Resynchronization Therapy in Dyssynchronous Canine HeartsCirculation: Arrhythmia and Electrophysiology, 2010
- Optimization of the Interventricular Delay in Cardiac Resynchronization Therapy Using the QRS WidthThe American Journal of Cardiology, 2009
- Left Ventricular Endocardial Pacing Improves Resynchronization Therapy in Canine Left Bundle-Branch HeartsCirculation: Arrhythmia and Electrophysiology, 2009
- The Optimized V‐V Interval Determined by Interventricular Conduction Times Versus Invasive Measurement by LVdP/dtMAXJournal of Cardiovascular Electrophysiology, 2008
- Electrocardiographic imaging of patients with heart failure with left bundle branch block and response to cardiac resynchronization therapyJournal of Electrocardiology, 2007
- Calculation of effective VV interval facilitates optimization of AV delay and VV interval in cardiac resynchronization therapyHeart Rhythm, 2007
- Tailoring cardiac resynchronization therapy using interventricular asynchrony. Validation of a simple modelAmerican Journal of Physiology-Heart and Circulatory Physiology, 2006
- Ischemic ventricular arrhythmias during heart failure: A canine model to replicate clinical eventsHeart Rhythm, 2005