Performance Characteristics of Long Axial Field-of-View PET Scanners With Axial Gaps
- 28 September 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 5 (3) , 322-330
Abstract
The introduction of long (>60 cm) axial field-of-view (LAFOV) PET systems has shown their potential for clinical and research applications. LAFOV scanners are expensive, so there is interest in designing systems with longer axial coverage while mitigating cost by introducing detector gaps. We used measurements on the PennPET Explorer (64-cm axial field-of-view (AFOV) prototype) and simulations of scanners up to 143-cm long to assess scanner performance with axial gaps introduced by varying the number of detector rows in each ring. Removing detectors reduces the total sensitivity and results in a nonuniform axial noise profile. Axial resolution shows small (<; 0.5 mm) loss from the edge of the AFOV to the center, even for a 143-cm AFOV scanner with an unrestricted acceptance angle. The presence of large axial gaps increases the variability in axial resolution and contrast recovery across the AFOV compared to a system without gaps. More modest axial gaps show less variable behavior. The results suggest that designs where the gap is no larger than one-half of the width of a detector ring may be preferred, although the optimal choice of scanner design with the tradeoffs of performance and AFOV will depend on its intended usage.Keywords
Funding Information
- National Institutes of Health (R01-CA113941, R01-CA206187, R01-CA225874)
- Research Agreement with Philips Healthcare
This publication has 35 references indexed in Scilit:
- Parallax error in long-axial field-of-view PET scanners—a simulation studyPhysics in Medicine & Biology, 2016
- Analytic TOF PET reconstruction algorithm within DIRECT data partitioning frameworkPhysics in Medicine & Biology, 2016
- Determination of Accuracy and Precision of Lesion Uptake Measurements in Human Subjects with Time-of-Flight PETJournal of Nuclear Medicine, 2014
- Development of a small prototype for a proof-of-concept of OpenPET imagingPhysics in Medicine & Biology, 2011
- Why is TOF PET reconstruction a more robust method in the presence of inconsistent data?Physics in Medicine & Biology, 2010
- A proposal of an open PET geometryPhysics in Medicine & Biology, 2008
- Iterative image reconstruction using geometrically ordered subsets with list-mode data2005
- GATE: a simulation toolkit for PET and SPECTPhysics in Medicine & Biology, 2004
- Performance of a Fast Maximum Likelihood Algorithm for Fully 3D PET Reconstruction1996
- Practical considerations for 3-D image reconstruction using spherically symmetric volume elementsIEEE Transactions on Medical Imaging, 1996